background image

 

 

Parallel Probe Resistivity Method 
 

The PRP method is used to give fast electrical resistivity 
measurements on flat homogeneous materials. Parallel 
probes are mounted on the back of the SRM500.  
 
Note: It may be used on multilayered materials, but this 
should be noted along with the temperature and humidity 
value on the data report. 
 
Make sure that surfaces to be tested are clean and free of 
contaminants. 
 
1.  Place the meter on the surface to be tested. 
2.  Move the switch to desired test voltage position, either 

10 or 100 volts. 

3.  Press and hold the test button. After approximately 15 

seconds, the meter will display the resistivity, humidity 
and temperature on the LCD screen.  

 
The resistivity reading will be in ohms per square, tem-
perature in centigrade, and relative humidity in percent. 
 
The meter will keep  updating the display  while the button 
is held down and will continue to display the last test read-
ing  for  approximately  45  seconds  after  the  button  is  re-
leased. 

 

 

Summary of Contents for Ohm Metrics SRM500K

Page 1: ...Instruction Manual Surface Resistance Testing Kit Model SRM500K Ohm Metrics...

Page 2: ...1 Ohm Metrics 2 2 Operation Power Option 2 Set Up 2 Parallel Probe Resistivity Testing 3 Probe Resistivity Testing 4 Resistance Between Two Points RTT 4 Surface to Ground Resistance RTG 5 3 Specificat...

Page 3: ...uickly measures resistance between two points RTT and surface to ground resistance RTG Suitable for factory au dits and test lab evaluations the Ohm Metrics SRM500K also measures ambient temperature a...

Page 4: ...ce Verification TR53 All Ohm Metrics products can be calibrated All Ohm Metrics test and measurement products are warranted for 1 full year Power Options The SRM500K can be powered by a 9 volt battery...

Page 5: ...ee of contaminants 1 Place the meter on the surface to be tested 2 Move the switch to desired test voltage position either 10 or 100 volts 3 Press and hold the test button After approximately 15 secon...

Page 6: ...in dependent of a ground point 1 Connect the plug ends of the test leads into the 3 5mm jacks of the meter 2 Connect the banana plugs of the test coil cords into the ends of the 5 pound probes 3 Place...

Page 7: ...to a known ground point 3 Position the probe on the surface to be tested in accor dance with the desired test procedure 4 Press the test button until the resistivity humidity and temperature test val...

Page 8: ...matic ranging Power Supply 9V Battery PP3 120 volt AC Adapter Unit automatically powers off when not in use Probes Two 5 pound disk probes Two 3 parallel surface resistivity probes Read Out LCD alpha...

Page 9: ...7 Specifications Continued Accuracy for 10 volt scale 10e3 10e8 10 RH 90 10e8 10e12 20 RH 90 Accuracy for 100 volt scale 10e3 10e8 10 RH 90 10e9 10e12 20 RH 60...

Page 10: ...the date of shipment Liability is limited to servicing after evalu ating repairing or replacing any product returned to Transforming Technologies The company does not war rant damage due to misuse neg...

Page 11: ...ect their prod ucts and processes from the many serious problems as sociated with static electricity Transforming Technologies offers a wide range of unique and outstanding products to detect protect...

Page 12: ...3719 King Rd Toledo OH 43617 P 419 841 9552 F 419 841 3241 info transforming technologies com...

Reviews: