background image

26

27

EMC GUIDANCE

EMC GUIDANCE

Table 2 

 

Guidance and manufacturer’s declaration – electromagnetic Immunity 

Immunity Test

 

±8 kV contact
±2 kV, ±4 kV, ±8 kV,  ±15 kV air 

±8 kV contact
±2 kV, ±4 kV, ±8 kV,  ±15 kV air 

 

±2 kV for power supply lines
±1 kV signal input/output 
100 kHz repetition frequency

±2 kV for power supply lines
Not Applicable
100 kHz repetition frequency

±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV, ±2 kV common mode    

±0.5 kV, ±1 kV differential mode
Not Applicable   

30 A/m
50 Hz / 60 Hz

30 A/m
50 Hz / 60 Hz

NOTE  U

is the a.c. mains voltage prior to application of the test level. 

Compliance level

Electrostatic
discharge (ESD) 
IEC 61000-4-2  

Power frequency 
magnetic field
IEC 61000-4-8

Voltage dips, short 
interruptions and 
voltage variations 
on power supply
 input lines 
IEC 61000-4-11

Electrical fast
transient/burst
IEC 61000-4-4   

Surge
IEC61000-4-5

IEC 60601-1-2 
Test level

0% U

T

; 0,5 cycle. At 0°, 45°, 90°, 135°,

180°, 225°, 270° and 315°.
0% U

T

; 1 cycle and 70% U

T

25/30 cycles; Single phase: at 0°.
0% U

T

; 250 / 300 cycle 

0% U

T

; 0,5 cycle. At 0°, 45°, 90°, 135°,

180°, 225°, 270° and 315°.
0% U

T

; 1 cycle and 70% U

T

25/30 cycles; Single phase: at 0°.
0% U

T

; 250 / 300 cycle 

Conduced RF
IEC61000-4-6

3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands 
between 0,15 MHz and 80 MHz
80% AM at 1 kHz

3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands 
between 0,15 MHz and 80 MHz
80% AM at 1 kHz

10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz

10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz

Radiated RF
IEC61000-4-3

Table 3

 

 

 

 

 

 

 

385

0.3

27

Service 

Modulation 

380-390

1.8

450

430-470

2

0.3

28

710

704-787

745

780

0.2

0.3

9

810

870

930

800-960

2

0.3

28

2

0.3

28

2

0.3

28

0.2

0.3

9

1720

1845

1970

1700-
1990

2450

2400-
2570

5240

5500

5785

5100-
5800

Guidance and manufacturer’s declaration - electromagnetic Immunity

Test 
Frequency
(MHz)

Radiated RF 
IEC61000-4-3 
(Test 
specifications 
for 
ENCLOSURE 
PORT 
IMMUNITY to 
RF wireless 
communicati- 
ons 
equipment) 

Band 
(MHz)

Distance 
(m)

IEC 
60601-1-2
Test Level
(V/m)

TETRA 
400

Pulse 
modulation  
18 Hz

GMRS 460,
FRS 460

FM  ± 5k Hz
deviation 
1 kHz sine

LTE Band
13,
17

Pulse 
modulation 
217 Hz

GSM 
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5

Pulse 
modulation 
18 Hz

GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1,
3, 4,25; 
UMTS

Pulse 
modulation 
217 Hz

Bluetooth,
WLAN,
802.11 
b/g/n, RFID
2450, LTE
Band 7

Pulse 
modulation 
217 Hz

WLAN
802.11
a/n

Pulse
modulation 
217 Hz

Compliance
 level 
(V/m) 

27

28

9

28

Maximum 
Power
(W)

28

9

28

Summary of Contents for TMB-1490-BS

Page 1: ...490 BS Please do read the user manual carefully and thoroughtly so as to ensure the safe usage of this product and keep the manual well for further reference in case you have problems version 1 0 Guan...

Page 2: ...ords Delete the Records INFORMATION FOR USER 18 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 20 What are systolic pressure and diastolic pressure What is the standard blood pressure classifi...

Page 3: ...rd and using Measurement Principle This product uses the Oscillometric Measuring method to detect blood pressure Before every measurement the unit establishes a zero pressure equivalent to the air pre...

Page 4: ...high frequency HF surgical equipment at the same time CAUTION Warning This device is not AP APG equipment Do not use the device where flammable anesthetic are present or in environments mixture with...

Page 5: ...t three records Bluetooth icon The bluetooth icon blinks when the bluetooth is working Pulse in beats per minute Pulse display Deflation symbol The cuff is deflating Indicate it is in the memory mode...

Page 6: ...t use different types of batteries together Do not dispose the batteries in fire Batteries may explode or leak Remove batteries if the device is not likely to be used for some time Worn batteries are...

Page 7: ...right year press SET to set down and turn to next step 4 Repeat steps 2 and 3 to set the MONTH and DAY 5 Repeat steps 2 and 3 to set the TIME FORMAT between 12h and 24h Pul min MEMORY TIME DATE S M ST...

Page 8: ...that the cuff is at the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep breaths Rest for 5 minutes before first measuring Wait at least 3 minutes between meas...

Page 9: ...nd it will finish the whole measurement save and transmit the measurement data for the desired user Take User A for example Install the APP and register an account Then set your personal information G...

Page 10: ...min MEMORY TIME DATE S M START STOP Recall the Records 2 Press the MEM or SET to get the record you want The most recent record 1 is shown first Each new measurement is assigned to the first 1 record...

Page 11: ...ase Avoid dusty and unstable temperature environment Do not attempt to clean the reusable cuff with water and never immerse the cuff in water Tips for Measurement Within 1 hour after dinner or drinkin...

Page 12: ...r Heartbeat Detector CAUTION The appearance of the IHB icon indicates that a pulse irregularity consistent with an irregular heart beat was detected during measurement Usually this is NOT a cause for...

Page 13: ...e measurement Relax for a moment and then measure again Relax for a moment and then measure again The measurement process does not detect the pulse signal Loosen the clothing on the arm and then measu...

Page 14: ...re operating normally Warning Use of accessories transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions...

Page 15: ...ngle phase at 0 0 UT 250 300 cycle Conduced RF IEC61000 4 6 3 V 0 15 MHz 80 MHz 6 V in ISM and amateur radio bands between 0 15 MHz and 80 MHz 80 AM at 1 kHz 3 V 0 15 MHz 80 MHz 6 V in ISM and amateur...

Page 16: ...led and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If t...