LCD display signal
CAUTION
* The service life of the cuff may vary by the frequency of washing, skin condition, and
storage state. The typical service life is 10000 times.
* It is recommended that the performance should be checked every 2 years and after
maintenance and repair, by retesting at least the requirements in limits of the error of the
cuff pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).
* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT
according to the local guidelines.
* Manufacturer will make available on request circuit diagrams, component part lists,
descriptions, calibration instructions, etc., to assist to service personnel in parts repair.
* The plug/adapter plug pins insulates the device from the main supply. Do not position the
device in a position where it is difficult to disconnect from the supply mains to safely
terminate operation of ME equipment.
* The operator shall not touch output of batteries/adapter and the patient simultaneously.
* Cleaning: Dust environment may affect the performance of the unit. Please use the soft
cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile
cleaners.
* The device doesn’t need to be calibrated within two years of reliable service.
* If you have any problems with this device, such as setting up, maintaining or using,
please contact the SERVICE PERSONNEL of Transtek. Don’t open or repair the device
by yourself in the event of malfunctions. The device must only be serviced, repaired and
opened by individuals at authorized sales/service centers.
* Please report to Transtek if any unexpected operation or events occur.
* Keep the unit out of reach of infants, young children or pets to avoid inhalation or
swallowing of small parts. It is dangerous or even fatal.
* Be careful to strangulation due to cables and hoses, particularly due to excessive length.
* At least 30 min required for ME equipment to warm from the minimum storage
temperature between uses until it is ready for intended use. At least 30 min required for
ME equipment to cool from the maximum storage temperature between uses until it is
ready for intended use.
* This equipment needs to be installed and put into service in accordance with the
information provided in the ACCOMPANYING DOCUMENTS.
* Wireless communications equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies can affect this
equipment and should be kept at least a distance away from the equipment. The
distance d is calculated by the MANUFACTURER from the 80MHz to 5.8 GHz column of
Table 4 and Table 9 of IEC 60601-1-2:2014, as appropriate.
* Please use ACCESSORIES and detachable parts specified/authorized by
MANUFACTURE. Otherwise, it may cause damage to the unit or danger to the
user/patients.
* There is no luer lock connectors are used in the construction of tubing, there is a
possibility that they might be inadvertently connected to intravascular fluid systems,
allowing air to be pumped into a blood vessel.
* Please use the device under the environment which was provided in the user manual.
Otherwise, the performance and lifetime of the device will be impacted and reduced.
* Adapter is specified as a part of ME EQUIPMENT.
SYMBOL DESCRIPTION
EXPLANATION
Irregular heartbeat detected during measurement.
Irregular heartbeat
Heartbeat
Heart rate decteted during measurement.
Systolic blood pressure
The high pressure measured.
Diastolic blood pressure The low pressure measured.
User ID
Current Time
Time and date (year/month/day; hour:minute)
Memory display
Indicate it is in the memory mode and which group of memory it is.
The bluetooth transfer icon blinks when the bluetooth is working.
Bluetooth transfer icon
Blood pressure level
Indicates the blood pressure level.
appears when the monitor is operated by User 1.
appears when the monitor is operated by User 2.
appears when the monitor is operated by guest User “G”.
NOTE: User 1 and User 2, each with 60 memory spaces;
User G, no momery space.
mmHg
KPa
Measurement Unit of the blood pressure.
(1 kPa=7.5 mmHg)
Measurement Unit of the blood pressure. (1 mmHg = 0.133 kPa)
Pulse/minute
Measurement Unit of Heart Rate.
Low ba�ery
Ba�eries are low and need to be replaced
Defla�on symbol
The cuff is defla�ng.
INTRODUCTION
INTRODUCTION
6