background image

LCD display signal

CAUTION

* The service life of the cuff may vary by the frequency of washing, skin condition, and 

storage state. The typical service life is 10000 times.

* It is recommended that the performance should be checked every 2 years and after 

maintenance and repair, by retesting at least the requirements in limits of the error of the 

cuff pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).

* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT 

according to the local guidelines.

* Manufacturer will make available on request circuit diagrams, component part lists, 

descriptions, calibration instructions, etc., to assist to service personnel in parts repair.

* The plug/adapter plug pins insulates the device from the main supply. Do not position the 

device in a position where it is difficult to disconnect from the supply mains to safely 

terminate operation of ME equipment.

* The operator shall not touch output of batteries/adapter and the patient simultaneously.

* Cleaning: Dust environment may affect the performance of the unit. Please use the soft 

cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile 

cleaners.

* The device doesn’t need to be calibrated within two years of reliable service.

* If you have any problems with this device, such as setting up, maintaining or using, 

please contact the SERVICE PERSONNEL of Transtek. Don’t open or repair the device 

by yourself in the event of malfunctions. The device must only be serviced, repaired and 

opened by individuals at authorized sales/service centers.

* Please report to Transtek if any unexpected operation or events occur.

* Keep the unit out of reach of infants, young children or pets to avoid inhalation or 

swallowing of small parts. It is dangerous or even fatal.

* Be careful to strangulation due to cables and hoses, particularly due to excessive length.

* At least 30 min required for ME equipment to warm from the minimum storage 

temperature between uses until it is ready for intended use. At least 30 min required for 

ME equipment to cool from the maximum storage temperature between uses until it is 

ready for intended use.

* This equipment needs to be installed and put into service in accordance with the 

information provided in the ACCOMPANYING DOCUMENTS.

* Wireless communications equipment such as wireless home network devices, mobile 

phones, cordless telephones and their base stations, walkie-talkies can affect this 

equipment and should be kept at least a distance away from the equipment. The 

distance d is calculated by the MANUFACTURER from the 80MHz to 5.8 GHz column of 

Table 4 and Table 9 of IEC 60601-1-2:2014, as appropriate.

* Please use ACCESSORIES and detachable parts specified/authorized by 

MANUFACTURE. Otherwise, it may cause damage to the unit or danger to the 

user/patients.

* There is no luer lock connectors are used in the construction of tubing, there is a 

possibility that they might be inadvertently connected to intravascular fluid systems, 

allowing air to be pumped into a blood vessel.

* Please use the device under the environment which was provided in the user manual. 

Otherwise, the performance and lifetime of the device will be impacted and reduced.

* Adapter is specified as a part of ME EQUIPMENT.

SYMBOL DESCRIPTION

EXPLANATION

Irregular heartbeat detected during measurement.

Irregular heartbeat

Heartbeat

Heart rate decteted during measurement.

Systolic blood pressure

The high pressure measured.

Diastolic blood pressure The low pressure measured. 

User ID

Current Time 

Time and date (year/month/day; hour:minute)

Memory display

Indicate it is in the memory mode and which group of memory it is.

 

The bluetooth transfer icon blinks when the bluetooth is working.

Bluetooth transfer icon

Blood pressure level

Indicates the blood pressure level.

appears when the monitor is operated by User 1.
appears when the monitor is operated by User 2.
appears when the monitor is operated by guest User “G”.

NOTE:  User 1 and User 2, each with 60 memory spaces;

User G, no momery space.

mmHg
KPa

Measurement Unit of the blood pressure. 

(1 kPa=7.5 mmHg)

Measurement Unit of the blood pressure. (1 mmHg = 0.133 kPa)

Pulse/minute

Measurement Unit of Heart Rate.

Low ba�ery

Ba�eries are low and need to be replaced

Defla�on symbol

The cuff is defla�ng.

INTRODUCTION

INTRODUCTION

6

7

Summary of Contents for TMB-1491-BHJ

Page 1: ...onitor Please read the user manual carefully and thoroughly so as to ensure the safe usage of this product Keep this manual for further reference in case any issues arise Version 1 0 Guangdong Transte...

Page 2: ...the Records Delete the Records INFORMATION FOR USER 21 Tips for measurement Maintenance ABOUT BLOOD PRESSURE 23 What are systolic pressure and diastolic pressure What is the standard blood pressure c...

Page 3: ...such as cardiac pacemakers de brillators Contraindications This product uses the Oscillometric Measuring method to detect blood pressure Before every measurement the unit establishes a zero pressure...

Page 4: ...f patient blood circulation When measurement please avoid compression or restriction of the connection tubing CAUTION The device cannot be used with HF surgical equipment at the same time The ACCOMPAN...

Page 5: ...equipment to cool from the maximum storage temperature between uses until it is ready for intended use This equipment needs to be installed and put into service in accordance with the information prov...

Page 6: ...the measurement 6V 1A Do not use new and used ba eries together Do not use di erent types of ba eries together Do not dispose the ba eries in re Ba eries may explode or leak Remove ba eries if the dev...

Page 7: ...EM bu on will increase the number by one in a cycling manner When you get the right year press SET bu on to con rm the entry The screen will then show a blinking number represen ng the MONTH It is imp...

Page 8: ...between user 1 user 2 and USER G guest 1 Press SET bu on to ensure your choice it will show User ID and donE and then turn o 2 7 A er the UNIT is set the LCD will display donE rst then display all th...

Page 9: ...ment Sensi ve people including pregnant women pre eclamp c and those who implanted medical electronic instruments should avoid using the unit whenever possible Keep the monitor at least 20 cen meters...

Page 10: ...eep breaths Rest for 5 minutes before rst measuring Wait at least 3 minutes between measurements This allows your blood circula on to recover The pa ent must relax as much as possible and do not move...

Page 11: ...1 is shown rst Each new measurement is assigned to the rst 1 record All other records are pushed back one digit e g 2 becomes 3 and so on and the last record 60 is dropped from the list 2 Each press M...

Page 12: ...can delete all results for the selected user by following steps Take User 1 for example If there is no record when you press MEM bu on to check the record the right display will be shown 3 Tips for Me...

Page 13: ...n indicates that a pulse irregularity consistent with an irregular heart beat was detected during measurement Usually this is NOT a cause for concern However if the symbol appears o en we recommend yo...

Page 14: ...during use please check the following points PROBLEM SYMPTOM CHECK THIS REMEDY No power Low batteries Error message Display will not light up Ba eries are exhausted Replace with new ba eries Insert th...

Page 15: ...ery Powered Mode Internally Powered ME Equipment AC Adapter Powered Mode Class II ME Equipment A temperature range of 5 C to 40 C A rela ve humidity range of 15 to 90 non condensing but not requiring...

Page 16: ...ses Bio compatibility ISO 10993 1 2018 Biological evaluation of medical devices Part 1 Evaluation and testing within a risk management process ISO 10993 5 2009 Biological evaluation of medical devices...

Page 17: ...anufacturer s declaration electromagnetic emissions RF emissions CISPR 11 RF emissions CISPR 11 Group 1 Class B Class A Comply Compliance Harmonic emissions IEC 61000 3 2 Voltage fluctuations flicker...

Page 18: ...3 Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communicati ons equipment Band MHz Modulation W Distance m IMMUNITY TEST LEVEL V m TETRA 400 Pulse modulation b 18 Hz GMRS 460 FRS 460...

Reviews: