User's Manual                                                         Rev B / February 2021 / M125 Helium Compressor

 

 

 

97-00020-000    

 

                                                                                                                                           15 

 

 

6

 

 

Operations 

6.1

 

Before Switching ON the System 

After the compressor unit and its load (cryopump, cold head, etc.) are installed and connected, check the helium gas 

pressure as indicated by the pressure gauge mounted on the rear panel of the compressor unit. Refer to Table 4-3, for the 

proper static pressure readings for the compressor. 

 

If the helium pressure needs to be adjusted, refer to Sections 8.4.1 and 8.4.2

 

for procedures to release helium gas in order 

to reduce the pressure or to fill the compressor with more helium gas to increase the pressure. 

 

Do not remove the top cover of the compressor unit. Doing so will disable the unit due to the built-in safety interlock 

mechanism. 

6.2

 

Normal Operation 

The load of the compressor can be powered through the power connectors located on the front panel of the compressor. 

To start operation of the compressor and its load, do the following: 

 

1.

 

Open the coolant supply (water-cooled compressor model only). 

2.

 

Switch on the main power source. 

3.

 

Press the ON button to start the compressor. Both the compressor and its load should start simultaneously. 

 

During operation, check the coolant flow rate (water-cooled compressor) and the operating pressure frequently. Refer to 

Table 4-3 for required coolant flow rate. If it is too slow, make sure any problems associated with water supply or water 

outlet are resolved. Refer to Table 4-3 for proper helium pressure level for the compressor unit. If the helium pressure is 

too low, switch off the compressor unit. It may be necessary to perform a helium "topping-up" maintenance procedure as 

described in Section 8.4.1. If pressure drop-off happens frequently, there may be a substantial leak in the helium circuit of 

the compressor. In this case, contact Trillium US Inc. customer service immediately (see Section 2.2). 

 

To shut down the compressor unit, press the OFF button on the front panel. After that, allow coolant to continue to 

circulate for at least 10 more minutes before shutting off flow. 

7

 

Troubleshooting 

7.1

 

Troubleshooting Activities 

Table 6-1 describes some problems that users might encounter while operating the Model 125 Cryopump Compressor and 

provides solutions to those problems. 

 

If a compressor problem still persists after performing the corrective actions described in this section, please contact 

Trillium US Inc. Technical Support for further assistance (see Section 2.2).  

 

Table 7-1: Trouble Shooting Procedures 

Problem 

Possible Cause 

Corrective Action 

The compressors On/Off 

switch (SW1) remains in 

the 

On position when switched 

on, but the pump does not 

run. 

1

 

No power is coming from the power 

source. 

2

 

Incorrect or disconnected wiring 

within the compressor 

 

1

 

Check service fuses, circuit breakers, 

and wiring associated with the power 

source. Repair as needed. 

2

 

Check the compressor against the wiring 

schematic. See 

Figure 5

. 

Summary of Contents for 91-00003-0LW

Page 1: ...97 00020 000 Trillium US Inc Model 125 Cryogenic Helium Compressor User s Manual Rev B February 2021...

Page 2: ...or infringements of patents or other rights of third parties This document is the property of Trillium US Inc and by furnishing this information Trillium US Inc grants no license expressed or implied...

Page 3: ...ngs 3 3 2 4 Heat Related Warnings 3 3 3 Operator Instructions 3 4 Introduction 4 4 1 General Information about the Model 125 Compressor 4 4 1 1 Model 125 Features 4 4 1 2 Overview of Model 125 Compres...

Page 4: ...9 8 6 Returning Equipment 19 List of Figures Figure 1 Flow Diagram for Water Cooled Model 125 compressor 6 Figure 2 Flow Diagram for Air Cooled Model 125 compressor 7 Figure 3 Model 125 Dimensions 9 F...

Page 5: ...evise document September 2010 1 1 Conversion to new format Update figures Change adsorber change interval June 2011 1 2 Correct minor typographical errors Add maximum inlet water pressure specificatio...

Page 6: ...nd helium compressors which are refurbished and fully warranted Technical Support Our support engineers will help determine if your cryopump system is operating correctly so that you can get your syst...

Page 7: ...efore beginning any work on the compressor unit the compressor needs to be switched off then isolated from the power supply Connect or disconnect the flex lines joining the compressor and its load cry...

Page 8: ...drawing helium gas from a high pressure gas cylinder Detaching the helium flex lines when the compressor load is at low temperature may cause the pressure to rise in the system beyond the permissible...

Page 9: ...and are charged with ultra high purity 99 999 helium gas The heat exchanger removes the heat generated from the process of compressing helium in the capsule The heat generated by the capsule must be r...

Page 10: ...opens to allow safe operation External relief valve Purpose and Safety Function Opens the helium gas circuit to atmosphere if the helium gas pressure exceeds 350 PSI 24 1 Bar 4 1 4 Operational Flow Th...

Page 11: ...User s Manual Rev B February 2021 M125 Helium Compressor 97 00020 000 6 Figure 1 Flow Diagram for Water Cooled Model 125 compressor...

Page 12: ...User s Manual Rev B February 2021 M125 Helium Compressor 97 00020 000 7 Figure 2 Flow Diagram for Air Cooled Model 125 compressor...

Page 13: ...idth 19 50 inches 495 mm Weight 150 lb 68 kg Helium Pressure Static 190 5 PSIG 13 1 35 Bar Operating 250 10 PSIG 17 24 69 Bar supply Interface Cold head power connector mates with ASC and CTI drive ca...

Page 14: ...W M125 Air cooled Low Voltage Standard Drive Circuit 91 00003 1LA Table 4 5 Optional Accessories and Replacement Parts Accessories Replacement Parts Part Number Adsorber Water cooled model 80034 Adsor...

Page 15: ...to process the insurance information in a timely manner Therefore it is highly recommended that shipping carton be opened and the unit inspected whether or not it will be put into operation right away...

Page 16: ...nit is switched off Never twist the helium flex lines during the installation process Before connecting the helium flex lines follow these steps 1 Identify the helium Return low pressure and Supply hi...

Page 17: ...Detach the coupling of the helium flex line from the helium charge vent valve 8 Close the helium gas regulator on the supply cylinder 9 Seal the helium gas charge vent valve on the compressor unit by...

Page 18: ...g of the frequency switch on the front panel matches the frequency of the power supply where the unit is being installed Failure to do so will result in performance degradation of the system Electrica...

Page 19: ...use 1 AMP Not Shown F1 F2 F3 F4 50081 6 Fuse 3 AMP Not Shown F5 F6 31 00012 018 7 Relay Start K1 50085 8 Meter Elapsed Time ETM1 50089 9 Connector 15 Pin Chassis J1 50087 10 Connector Housing Mate N L...

Page 20: ...e sure any problems associated with water supply or water outlet are resolved Refer to Table 4 3 for proper helium pressure level for the compressor unit If the helium pressure is too low switch off t...

Page 21: ...5 Low power source voltage 6 Mechanical seizure 1 Confirm that sufficient cooling water for water cooled model is flowing to the compressor For air cooled model provide additional cooling to the surr...

Page 22: ...system In severe cases of such oil contamination the load cryopump cold head etc could cease to function completely To remove and replace the compressor adsorber follow the steps described in Sections...

Page 23: ...is depressed 8 Make sure the supply pressure gauge reads 190 5 PSIG 13 1 35 Bar If the pressure is either too high or too low follow the instructions in Section 8 4 1 to fill the helium gas to the pr...

Page 24: ...ge rises to 190 5 PSIG 13 1 35 Bar tightly close the charge valve 4 Insure that the helium charge valve on the compressor is tightly closed Then shut off the helium pressure regulator on the helium bo...

Reviews: