background image

User's Manual                                                         Rev B / February 2021 / M125 Helium Compressor

 

 

 

97-00020-000    

 

                                                                                                                                           17 

 

 

Caution: 

Loosening or detaching helium flex lines with the compressor load at low temperature without proper 

warming-up can result in loss of helium and/or pressure rise in the cryopump unit beyond its designed 

maximum pressure level. 

 

When transporting the compressor unit, follow these guidelines: 

 

 

Make sure the appropriate protective caps are properly secured before shipping. 

 

Always store the compressor unit in a dry place. Refer to Table 4-3 for proper storage environment. 

 

If a freezing temperature environment is anticipated whether during shipping or under storage, make certain the 

coolant in the compressor circuit is properly drained. 

 

Caution:  

The compressor unit should never be tilted more than 45 degrees either during shipping or in storage. 

8.3

 

Scheduled Preventative Maintenance Activity 

The only scheduled maintenance required on the Model 125 compressor is the replacement of compressor adsorber after 

every 15,000 hours of operation as indicated on the compressor elapsed time meter or every three years (whichever comes 

first). When the compressor is used with 50 Hz power, the actual elapsed time will be 1.2 times of that shown on the 

meter. 

 

The adsorber is used to keep the oil vapor out of the helium gas in the flow circuit of compressor unit and its load. After 

about 15,000 hours of operation (or every three years, whichever comes first), the effectiveness of the adsorber will 

decrease. It will then need to be replaced. Otherwise, the oil particles could accumulate on the cold surface of the 

compressor load, reducing the cooling performance of the overall system. In severe cases of such oil contamination, the 

load (cryopump, cold head, etc.) could cease to function completely. 

 

To remove and replace the compressor adsorber, follow the steps described in Sections 8.3.1 and 8.3.2

 

 

Caution: 

 Use only Trillium US Inc. supplied adsorber for replacement. Refer to Section 4.3 for part number. 

8.3.1

 

Remove the Compressor Adsorber 

To remove the compressor adsorber: 

1.

 

Turn off the compressor. 

2.

 

Use the two wrenches supplied with the Installation Kit to avoid loosening the body of the coupling from its 

adapter. Hold one wrench tight on the coupling half attached to the rear side of the compressor. Use the other 

wrench to loosen the coupling to the helium supply line. 

3.

 

Unscrew the two self-sealing coupling halves quickly to minimize minor gas leakage. 

4.

 

 contains an illustration of the self-sealing couplings. 

5.

 

Remove the top cover. 

6.

 

Remove the adsorber from the compressor as shown in Section 8.3.1. Save all nuts, bolts, and washers for 

installing the replacement adsorber. 

7.

 

The removed adsorber can be returned to Trillium US Inc. for credit. Section 2.2 provides the contact information. 

 

8.3.2

 

Install Replacement Adsorber 

To install the replacement adsorber: 

1.

 

Remove the dust caps from the self-sealing coupling halves at each end of the replacement adsorber. 

2.

 

Check the self-sealing connector flat rubber gasket to make sure that it is clean and properly positioned. 

3.

 

Place the adsorber back in the compressor using the nuts, bolts, and washers set aside during the removal process 

described in Section 8.3.1. 

4.

 

Install the two self-sealing coupling halves quickly to minimize minor gas leakage. 

5.

 

Use the two wrenches supplied with the Installation Kit, holding one wrench tight on the coupling half attached to 

the rear side of the compressor. Use the other wrench to tighten the coupling to the helium supply line. 

Summary of Contents for 91-00003-0LW

Page 1: ...97 00020 000 Trillium US Inc Model 125 Cryogenic Helium Compressor User s Manual Rev B February 2021...

Page 2: ...or infringements of patents or other rights of third parties This document is the property of Trillium US Inc and by furnishing this information Trillium US Inc grants no license expressed or implied...

Page 3: ...ngs 3 3 2 4 Heat Related Warnings 3 3 3 Operator Instructions 3 4 Introduction 4 4 1 General Information about the Model 125 Compressor 4 4 1 1 Model 125 Features 4 4 1 2 Overview of Model 125 Compres...

Page 4: ...9 8 6 Returning Equipment 19 List of Figures Figure 1 Flow Diagram for Water Cooled Model 125 compressor 6 Figure 2 Flow Diagram for Air Cooled Model 125 compressor 7 Figure 3 Model 125 Dimensions 9 F...

Page 5: ...evise document September 2010 1 1 Conversion to new format Update figures Change adsorber change interval June 2011 1 2 Correct minor typographical errors Add maximum inlet water pressure specificatio...

Page 6: ...nd helium compressors which are refurbished and fully warranted Technical Support Our support engineers will help determine if your cryopump system is operating correctly so that you can get your syst...

Page 7: ...efore beginning any work on the compressor unit the compressor needs to be switched off then isolated from the power supply Connect or disconnect the flex lines joining the compressor and its load cry...

Page 8: ...drawing helium gas from a high pressure gas cylinder Detaching the helium flex lines when the compressor load is at low temperature may cause the pressure to rise in the system beyond the permissible...

Page 9: ...and are charged with ultra high purity 99 999 helium gas The heat exchanger removes the heat generated from the process of compressing helium in the capsule The heat generated by the capsule must be r...

Page 10: ...opens to allow safe operation External relief valve Purpose and Safety Function Opens the helium gas circuit to atmosphere if the helium gas pressure exceeds 350 PSI 24 1 Bar 4 1 4 Operational Flow Th...

Page 11: ...User s Manual Rev B February 2021 M125 Helium Compressor 97 00020 000 6 Figure 1 Flow Diagram for Water Cooled Model 125 compressor...

Page 12: ...User s Manual Rev B February 2021 M125 Helium Compressor 97 00020 000 7 Figure 2 Flow Diagram for Air Cooled Model 125 compressor...

Page 13: ...idth 19 50 inches 495 mm Weight 150 lb 68 kg Helium Pressure Static 190 5 PSIG 13 1 35 Bar Operating 250 10 PSIG 17 24 69 Bar supply Interface Cold head power connector mates with ASC and CTI drive ca...

Page 14: ...W M125 Air cooled Low Voltage Standard Drive Circuit 91 00003 1LA Table 4 5 Optional Accessories and Replacement Parts Accessories Replacement Parts Part Number Adsorber Water cooled model 80034 Adsor...

Page 15: ...to process the insurance information in a timely manner Therefore it is highly recommended that shipping carton be opened and the unit inspected whether or not it will be put into operation right away...

Page 16: ...nit is switched off Never twist the helium flex lines during the installation process Before connecting the helium flex lines follow these steps 1 Identify the helium Return low pressure and Supply hi...

Page 17: ...Detach the coupling of the helium flex line from the helium charge vent valve 8 Close the helium gas regulator on the supply cylinder 9 Seal the helium gas charge vent valve on the compressor unit by...

Page 18: ...g of the frequency switch on the front panel matches the frequency of the power supply where the unit is being installed Failure to do so will result in performance degradation of the system Electrica...

Page 19: ...use 1 AMP Not Shown F1 F2 F3 F4 50081 6 Fuse 3 AMP Not Shown F5 F6 31 00012 018 7 Relay Start K1 50085 8 Meter Elapsed Time ETM1 50089 9 Connector 15 Pin Chassis J1 50087 10 Connector Housing Mate N L...

Page 20: ...e sure any problems associated with water supply or water outlet are resolved Refer to Table 4 3 for proper helium pressure level for the compressor unit If the helium pressure is too low switch off t...

Page 21: ...5 Low power source voltage 6 Mechanical seizure 1 Confirm that sufficient cooling water for water cooled model is flowing to the compressor For air cooled model provide additional cooling to the surr...

Page 22: ...system In severe cases of such oil contamination the load cryopump cold head etc could cease to function completely To remove and replace the compressor adsorber follow the steps described in Sections...

Page 23: ...is depressed 8 Make sure the supply pressure gauge reads 190 5 PSIG 13 1 35 Bar If the pressure is either too high or too low follow the instructions in Section 8 4 1 to fill the helium gas to the pr...

Page 24: ...ge rises to 190 5 PSIG 13 1 35 Bar tightly close the charge valve 4 Insure that the helium charge valve on the compressor is tightly closed Then shut off the helium pressure regulator on the helium bo...