Ac
cessor
ies
7 Technical Data
7.1 Technical Specifications
Measurement principle
Luminescence
Parameter
Dissolved Oxygen
Measurement range
0...20 mg/L
0...20 ppm
0...200 %
Measurement accuracy
± 0.1 mg/L
± 0.1 ppm
± 1 %
Resolution
0.01
Response time
90 % of the value in less than 60 seconds
Measurement interval
> 5 s
Inflow Velocity
no movement necessary
Temperature Compensation
Via NTC (compensation active for temperature lower than 0 °C)
Measurement range
(temperature)
0...+50 °C
Resolution (temperature)
0.01 °C
Accuracy (temperature)
0.5 °C
Membrane Cap
No cross-sensitivity with : pH 1 – 14 ; CO2, H2S, SO2
Cross-sensitivity to Organic solvents, such as acetone, toluene, chloroform or methylene
chloride Chlorine gas
Material
Standard Version in passivated Stainless steel 316L body, strainer and screw,
For Seawater application Version in Titanium body, strainer and screw,
Cable : polyurethane jacket
Steam gland : Polyamide
Patch with active material (black) – Membrane : Optical isolation silicon
Dimensions (L x Ø)
146 mm x 25 mm
~ 5.7˝ x 1˝
Weight
stainless steel
~ 450 g
~ 1 lbs
titanium
~ 300 g
~ 0.7 lbs
Interface
RS-485 (Modbus RTU)
Power consumption
1 W
Power supply
12 V (± 10 %)
Sensor cable
2 m and 10
m