D01-915en201911 Manual eCHEM Chlorine Dioxide
Chlorine Dioxide
//
Commissioning
Place the membrane cap on a clean surface and fill to the rim
with gel electrolyte.
Place the G-holder on a clean surface with the fabric facing
down and moisten the G-holder with gel electrolyte.
Use the tweezers [1] to grasp the G-holder [2] with the fabric
facing down and place the G-holder centrally on the memb
-
rane in the membrane cap [3].
Carefully pull the tweezers out of the membrane cap again.
The G-holder is now placed in the membrane cap.
Place the sensor body [1] upright on the membrane cap [2]
and turn it counterclockwise until the thread is engaged.
Slowly screw the sensor body into the membrane cap.
Make sure that the membrane cap is completely screwed onto the sensor body, i.e. as far as it will go.
Neither touch nor strike the membrane!
The sensor is ready for commissioning.
1
2
3
1
2
Rinse off electrolyte residues adhering to the outside of the
sensor with tap water.
NOTICE