Technical Data
7 Technical Data
7.1 Technical Specifications
Measure-
ment
technol-
ogy
Light source
LED
Detector
Photodiode
Measurement principle
Fluorescence
Parameters
see parameter list, chapter 7.2
Measurement range
0...200 µg/L or 0...500 µg/L
Measurement accuracy
± 5 %
Turbidity compensation
no
Data logger
no
Reaction time T100
6 s
Measurement interval
3 s
Housing material
Stainless steel (1.4571/1.4404) or titanium (3.7035)
Dimensions (L x Ø)
171 mm x 36 mm
Weight
stainless steel
0,5 kg
titanium
0,4 kg
Interface
digital
Ethernet (TCP/IP)
RS-232 or RS-485 (Modbus RTU)
Power
con-
sumption
typical
< 1 W
with network
< 1.6 W
Power supply
12...24 VDC (± 10 %)
Required supervision
typically ≤ 0,5 hours per month
Calibration/maintenance
interval
24 months
System compatibility
Modbus RTU
Warranty
1 year (EU: 2 years)
INSTALLATION
Max.
pressure
with Subconn
30 bars
with fixed cable
3 bars
in flow cell
1 bar, 2...4 L/min
Protection type
IP68
Sample temperature
+2...+40 °C
Ambient temperature
+2...+40 °C
Storage temperature
-20...+80 °C
Inflow velocity
0.1...10 m/s