Fig. 18. Web interface: Measurement sub-item Settings
In the sub-item settings page (Fig. 18), you can enter adjustments by clicking on the "Edit"
button. Comments entered in the comments field are combined with measured values and
spectra. You may also enable automatic measurements and specify the corresponding
measurement interval.
Under “Spectra Type” you can select the spectrum type that will be sent for example to a
TriOS
controller or Modbus communication device. You can choose between attenuation,
transmission and calibrated. If you select calibrated, the other spectra types will not be
calculated. If you choose transmission, you get calibrated spectra, too, but no attenuation
spectra.
Only if you activate attenuation, all 3 spectra types will be saved to the
internal data logger.