5.2 Measurement Properties
5.2.1 Influences on the Measurement
•
The pH measurement depends on the following parameters:
•
High temperatures of the measuring medium accelerate the ageing of the electrode.
•
Note the immersion depth: a too shallow immersion depth can lead to fluctuating measured values.
•
Air bubbles on the sensor can cause measurement errors.
•
Alkali error: at pH values above pH12, alkali ions (Li+, Na+) can cause lower pH values to be displayed, as the
alkaliions are detected in addition to the H+ ions.
•
Flow can influence the measured value.
5.2.2 Temperature Compensation
Temperature compensation is performed automatically by the integrated temperature sensor (Pt1000).