38
D01-051en202111 Manual TriBox3
Advanced
Use
Malfunction &
Maintenance
Commis-
sioning
Introduction
General
Information
FA
Q
Technical Data
W
arranty
Customer
Service
Contact
Keyword
Index
Accessories
Use
Use
//
TriBox3
The subitem “Cleaning” (menu item “Options”) allows the user to set the cleaning intervals.
To make use of an automatic cleaning interval, this must first be activated by checking the box and setting a
cleaning interval. The cleaning time and the pause before the measurement can also be defined here.
Cleaning can also be triggered and defined whether the valve and the relay are activated during cleaning.
When the valve switches, the air must come out of the outlet and stop completely
when the switching procedure has finished. If air is not released or if the valve
does not close correctly at the end of the switching procedure, both hoses must
be swapped.
NOTICE
The compressed air valve must never be exposed to pressure greater than 0.7 Mpa
(7 bars). Otherwise, the valve can be damaged.
NOTICE
To ensure that settings made are not lost when reconfiguring, it is essential to save a restore
point beforehand.
Please note that the measurement and cleaning intervals should be matched to each other
to avoid suspension of measurements. If this is necessary, use the recommended settings
described in the table below.
Summary of Contents for TriBox3
Page 1: ...TriBox3 OPERATING INSTRUCTIONS...
Page 2: ......