6.2.4 Checking the Relay
6.2.3 Checking the Valve
To check the functioning of the valve, the valve can be directly controlled via the menu. The compressed air
from the input should then be at the output (below or closer to the wall). A clicking sound from the valve should
be heard clearly every time it switches (see chapter 4.5).
To check the functioning of the relay, it can be directly controlled via the menu. A clicking sound from the relay
should be heard clearly every time it switches. The transit can be measured at the corresponding inputs and
outputs (see chapter 5.3).
The internal valve may only be used in unpressurized media. In the event of water
ingress through the valve, no warranty claims can be made against TriOS.
Danger of electric shock and fire. Only qualified personnel should carry out
the inspection check described in this chapter of the operating instructions.
DANGER
NOTICE