Triplett Model 1101-B
Cat. No.: 1101
Made in China
Triplett Three Year Limited Warranty
Triplett / Jewell Instruments warrants instruments and test equipment manufactured by it to be free from
defective material or workmanship and agrees to repair or replace such products which, under normal
use and service, disclose the defect to be the fault of our manufacturing, with no charge within three
years of the date of original purchase for parts and labor. If we are unable to repair or replace the
product, we will make a refund of the purchase price. Consult the Instruction Manual for instructions
regarding the proper use and servicing of instruments and test equipment. Our obligation under this
warranty is limited to repairing, replacing, or making refund on any instrument or test equipment which
proves to be defective within three years from the date of original purchase.
This warranty does not apply to any of our products which have been repaired or altered by unauthorized
persons in any way so as, in our sole judgment, to injure their stability or reliability, or which have been
subject to misuse, abuse, misapplication, negligence, accident or which have had the serial numbers
altered, defaced, or removed. Accessories, including batteries and fuses, not of our manufacture used
with this product are not covered by this warranty.
To register a claim under the provisions of this warranty, contact Triplett / Jewell Instruments Customer
Service Department for a Return Authorization Number (RMA) and return instructions. No returned
product will be accepted without an RMA number. Upon our inspection of the product, we will advise
you as to the disposition of your claim.
ALL WARRANTIES IMPLIED BY LAW ARE HEREBY LIMITED TO A PERIOD OF THREE YEARS FROM
DATE OF PURCHASE, AND THE PROVISIONS OF THE WARRANTY ARE EXPRESSLY IN LIEU OF
ANY OTHER WARRANTIES EXPRESSED OR IMPLIED.
The purchaser agrees to assume all liability for any damages and bodily injury which may result from the
use or misuse of the product by the purchaser, his employees, or others, and the remedies provided for
in this warranty are expressly in lieu of any other liability Triplett / Jewell Instruments may have, including
incidental or consequential damages.
Some states (USA ONLY) do not allow the exclusion or limitation of incidental or consequential damages,
so the above limitation or exclusion may not apply to you. No representative of Triplett / Jewell Instruments
or any other person is authorized to extend the liability of Triplett / Jewell Instruments in connection with
the sale of its products beyond the terms hereof.
Triplett / Jewell Instruments reserves the right to discontinue models at any time, or change specifications,
price or design, without notice and without incurring any obligation.
This warranty gives you specific legal rights, and you may have other rights which vary from state to
state.
TRIPLETT PRODUCT RETURN INSTRUCTIONS
In the unlikely event that you must return your 1101-B for repair,the following steps must be taken.
1) Call 1-800-TRIPLETT to obtain a Return Material Authorization (RMA) number from
Customer Service.
2) Enclose a copy of the original sales receipt showing date of purchase.
3) Print the RMA number on the outside of the shipping container.
4) Return to: Triplett / Jewell Instruments
850 Perimeter Road
Manchester, NH 03103
ATTN: Repair Dept.
Be sure to include a full description of the
problem, and a telephone number, street
address, or email address, where you can be
contacted, and a return address where the
meter can be shipped to upon repair.
84-884
3/10
7-2 Measuring DC Voltage
7-2-1
Connect black test lead to COM jack and red lead to V
Ω
mA jack.
7-2-2
Set the Rotary Switch to the desired
V
position.
7-2-3
Observing polarity, connect test leads across the circuit to be measured and read voltage
from LCD.
NOTE:
1.
When the quantity to be measured is unknown, set the Range Switch to its highest DC range
(i.e. 600V) before connecting test leads to circuit.
2.
If ‘1’ or ‘-1’ is displayed, it indicates ‘overrange’ . Either select a higher range or do not
measure.
3.
Although the 1101-B may measure voltages higher than 600V, the 1101-B may be damaged,
or may pose a shock hazard or other injury.
4.
Use caution when measuring high voltages.
7-3 Measuring AC Voltage
7-3-1
Connect black test lead to COM jack and red lead to V
Ω
mA jack.
7-3-2
Set the Rotary Switch to the desired V~ position.
7-3-3
Observing polarity, connect test leads across the circuit to be measured and read voltage
from LCD.
NOTE:
1.
When the quantity to be measured is unknown, set the Range Switch to its highest AC range
(i.e. 600V) before connecting test leads to circuit.
2.
If ‘1’ or ‘-1’ is displayed, it indicates ‘overrange’ . Either select a higher range or do not
measure.
3.
Although the 1101-B may measure voltages higher than 600V, the 1101-B may be damaged,
or may pose a shock hazard or other injury.
4.
Use caution when measuring high voltages.
7-4 Measuring DC Current
7-4-1
Connect black test lead to COM jack and the red lead to the V
Ω
mA jack for a maximum
200mA current. For a maximum 10A current, move the red lead to the 10A jack.
7-4-2
Set the Rotary Switch to the desired
A
range position.
7-4-3
Connect test leads in series with the load under measurement.
7-4-4
Read current from LCD.
NOTE:
1.
When the quantity to be measured is unknown, set the Range Switch to its highest range
(i.e. 200mA or 10A depending on the jack used.) before connecting test leads to circuit.
2.
If ‘1’ or ‘-1’ is displayed, it indicates ‘overrange’. Either select a higher range or do not
measure.
3.
Current above 200mA on the V
Ω
mA jack with blow the fuse. The 10A jack is not fused. Do
not apply more than 10A. Do not measure 10A for longer than 10 seconds. Allow a 15 minute
cooldown.
4.
If fuse blows, replace by removing battery cover and battery, stretch red ‘boot’ to remove,
remove case screw at bottom, open case.
7-5 Measuring Resistance
7-5-1
Connect the black test lead to COM jack and the red lead to the V
Ω
mA jack.
7-5-2
Set the rotary switch to the desired
Ω
range position.
7-5-3
Connect test leads across the resistance under measurement.
7-5-4
Read the resistance on the LCD.
NOTE:
1.
DO NOT APPLY VOLTAGE WHEN MEASURING RESISTANCE.
2.
If ‘1’ or ‘-1’ is displayed, it indicates ‘overrange’ .This means the test leads are not connected
to the circuit, or the resistance exceeds the measurement ability of the selected range. Select a
higher range to obtain a measurement.
3.
When checking in-circuit resistance, be sure the circuit under test has all power removed
and that all capacitors have been discharged fully.
7-6 Measuring Temperature
7-6-1
Set the rotary switch to the
°
F position.
7-6-2
The LCD will show the current temperature of the 1101-B.
7-6-3
To use the thermocouple temperature probe, insert the red plug of the thermocouple
into the V
Ω
mA jack, and black plug into the COM jack.
7-6-4
The temperature of the thermocouple tip is displayed on the LCD.
7-6-5
The probe included with the 1101-B is useful up to 300
°
F. To measure higher tempera-
tures, use a high temperature probe.
7-7 Transistor Test
7-7-1
Set the rotary switch to the ’hFE’ position.
7-7-2
Determine whether the transistor under testing is NPN or PNP and locate the emitter,
base and collector leads. Insert the leads into proper holes of the hFE socket on the front panel.
7-7-3
Read the approximate hFE value on the LCD.
7-8 Diode Test
7-8-1
Connect the black test lead to COM jack and the red test lead to V
Ω
mA jack. (the polarity
of red lead is ‘+’)
7-8-2
Set the rotary switch to the
range position.
7-8-3
Connect the red lead to the anode and the black lead to the cathode of the diode under
test.
7-8-4
Read the diode forward voltage on the LCD.
NOTE:
1.
The meter will show approximate forward voltage drop of the diode.
2.
If the leads are reversed or not connected , ‘1’ should be displayed.
7-9 Continuity Test
7-9-1
Connect the black test lead to COM jack and the red test lead to V
Ω
mA jack.
7-9-2
Set the rotary switch to the
position.
7-9-3
Connect test leads across two points of the circuit under test.
7-9-4
If continuity exists (i.e. resistance less than about 50
Ω
), the built-in beeper will sound.
NOTE:
If the leads are connected to an ‘open circuit’, ‘1’ will be displayed.
8. Maintenance
8-1
Before attempting to remove the battery door or open the case, be sure that test leads have
been disconnected from live circuitry to avoid electric shock hazard.
8-2
To avoid electrical shock or injury, remove test leads from live circuitry before replacing the
fuse. For protection against fire or injury, replace fuse only with approved 200mA/500V fuse.
8-3
If the test leads are damaged or any wire is exposed, replace the leads before using the
meter.
8-4
Do not use harsh chemicals or solvents to clean the meter.
8-5
Do not use the meter with the battery cover or back cover removed. User injury may result.
8-6
If the meter behaves abnormally, discontinue usage and have the meter serviced or re-
placed.
8-7
Remove battery if the meter will not be used for more than 1 month.
9. Accessories
1.
Test Leads: Triplett 79-760
3.
‘K’ type Thermocouple: Triplett 79-795
2.
Fuse: Fast 200mA/500V : Triplett 3207-135
4.
Instruction Manual: Triplett 84-884
www.calcert.com
sales@calcert.com
1.888.610.7664
0
5
10
15
20
25
30