Operating Instructions
RANGE Button
The Auto Range mode automatically selects the proper range for
the measurement being made and is generally the best mode for
most applications. For measurement situations requiring that a
range be manually selected, perform the following:
1. Momentarily press the
RANGE
button. The “
AUTO
”
indicator will no longer be shown on the LCD display.
2. Momentarily press the
RANGE
button to step through
the available ranges until the desired range is selected.
3. To exit the Manual Ranging mode, press and hold the
RANGE
button until the “
AUTO
” indicator reappears.
NOTE:
The range button does not work on Frequency,
Duty Cycle, Capacitance
or Temperature.
MODE Button
Used to select AC or DC, Frequency or Duty Cycle,
Resistance, Continuity or Diode Test, and °C or °F.
REL button
The RELATIVE function zeros out the reading on the display and
stores it as a reference. Subsequent readings will be displayed as
the relative difference between the actual measurement and the
stored reference value. To activate, press the REL button. The
“REL” indicator will appear on the LCD display along with the
relative reading. Press the REL button again to return to normal
operation.
NOTE:
The meter does not Auto Range when the Relative mode is
active. The
display will read OL if the difference exceeds the range.
When this occurs, exit REL and use the
RANGE
button to select a
higher range. REL does not work on Frequency, Duty Cycle,
Temperature, Continuity or Diode Test.