8.2 Cautions during Measurement
●Good measurements are made only when stable
coupling symbol appears;
●When there is a large amount of coupling agent on
the object surface, it will cause mismeasurement.
Therefore, at the end of the measurement, the
probe should be quickly removed from the surface.
●The probe surface is acrylic resin and should be
gently pressed when used. When measuring rough
surface, try to reduce the probe scratching on it.
●When measuring at room temperature, the surface
of the measured object should not exceed 60
℃
,
otherwise the probe can no longer be used.
●If the probe wears, the indicating value will be
unstable. Please replace it.
8.3 Clean the Standard Thickness Block
The standard thickness block should avoid rust.
Wipe the block after use, and don't come into
contact with sweat. If not used for a long time,
a little grease should be coated on the surface of
the block, and the grease can be wiped after the
next use.