Default number of discrete SloScans to perform.
QuikScan_Speed=30
Default interval (ms) between points for a continuous QuikScan.
QuikScan_InterpolationFactor=100
Default number of points to interpolate when generating a set of points for a QuikScan.
QuikScan_SmoothingFactor=50
Default smoothing of QuikScan points, to avoid sharp turns during a serpentine track.
QuikScan_AccelDecelPoints=32
Default duration of acceleration/deceleration of a continuous QuikScan.
QuikScan_AccelDecelCurve=0.4
Default magnitude of acceleration/deceleration of a continuous QuikScan.
[Excitation]
AO_Port=Dev1/ao0
Specifies port to be used for output of an excitation signal. The available ports are listed
under NI-MAX.
Channel=1
Specifies the channel number to assign to the excitation channel. This number must be
NumSQUIDs <= n <= (NumChans-1), and it must be after the
Calibration_(T/V)=8.47e-8
Specifies the calibration of the excitation coil. If = 0, then the coil parameters listed below
are used to calculate an approximate calibration.
Frequency_(Hz)=15.00
Default frequency of the excitation signal. Allowed values are in the range 0 to 1000 Hz.
Amplitude_(T)=1e-9
Default amplitude of the excitation signal.
Samples_Per_Buffer_(N)=250
Points in the excitation buffer.
Cycles_Per_Buffer_(N)=5
Periodic cycles in the excitation buffer.
Offset_(V)=0.00
Default amplitude of the excitation signal. Allowed values are in the range -10 to 10 V.
Type=SineWave