24
P5596.1600 rev B
4.5.1 Calibrate
This enables you to
Calibrate
the
Sentro
Wireless
.
From the
Main Menu
press
L,
navigate to
Calibrate
and press
R
to enter the
Calibrate
Menu
.
4.5.1.1 Calibrate
Calibrate the
Zero
and
Span
values of the
Sentro Wireless
using clean air and test gas
of a known concentration.
Checkpoint
Ensure the area where the
Sentro Wireless
is being calibrated is well ventilated.
Observe applicable local Health and Safety
legislation and applicable local procedures
when handling test gases.
Use a cross head screwdriver and remove the
four screws securing the front cover and
swing it out of the way. Fit a gas hood to the
front of the
Sentro eModule
, the gas hood is a
push fitting.
Using suitable tubing connect a cylinder of
clean air test gas to the gas hood fitted to the
Sentro eModule
.
Open the valve and apply clean air at a rate
of 0.5 litre/min. Wait for the reading to settle
at a stable value.
Press
R
to
Zero
the displayed value. Wait for
the reading to stabilise at zero.
Summary of Contents for Sentro Wireless TX6355
Page 1: ...TX6355 Sentro Wireless...
Page 2: ......
Page 6: ...6 P5596 1600 rev B www trolex com 1 4 Dimensions...
Page 10: ...10 P5596 1600 rev B www trolex com...
Page 11: ...P5596 1600 rev B 11 www trolex com TX6355 User Manual...
Page 18: ...18 P5596 1600 rev B www trolex com 4 Setup and Calibration 4 1 Controls and Indicators...
Page 19: ...P5596 1600 rev B 19 www trolex com TX6355 User Manual...
Page 20: ...20 P5596 1600 rev B www trolex com 4 2 Software Menus...
Page 56: ...56 P5596 1600 rev B www trolex com...
Page 57: ...P5596 1600 rev B 57 www trolex com TX6355 User Manual...