P5596.1600 rev B
45
TX6355 User Manual
5. Diagnostics and
Maintenance
5.1 Diagnostic Messages
Sensor Over-range
If the
Sentro Wireless
detects gas levels above
the safe working limit of the sensing element:
Pellistor Gas Sensor
•
The transmitted output signal from the
sensor will be clamped at full scale to prevent
ambiguous output data from being transmitted
and will latch in this condition
•
The pellistor in the gas sensing module will
be switched into a protect state
to prevent oxidisation damage. The pellistor
will automatically reset after 600 seconds and
will latch in this state
•
Check that the gas concentration has
receded, then reset the eModule by removing
it for 2 seconds
Electrochemical, Infrared or LED Infrared
Gas Sensor
•
The screen will display the error message