TX6351 • TX6352 User Manual
P5536.1600 rev A 11
TX6352
2 wire. Remote powered
4 to 20 mA analogue output
Supply voltage:
10 to 14 V
Max. line load:
90 R
Type of sensor:
Toxic only
1.7 Sentro eModules
Plug-in pre-calibrated modules with standardised output data.
• Each module stores all the necessary data about its type identification, sensing
range and specific calibration. This data is automatically recognised by Sentro
when the module is loaded into the module bay.
• The modules are pre-calibrated so can be substituted at any time by a
replacement module – usually of the same type, but an alternative may be
inserted if required.
• Insert coding stops to prevent invalid module combinations.
• The sensing module will identify itself when plugged into the sensor housing
and auto configuration will take place.
• All Sentro Modules have two output alarm signals for GENERAL alarm and
HIGH alarm. Default values are entered during manufacture and these can be
changed to preferred values.
• The alarm signals can be set to illuminate built-in flashing LED indicators.
• The signals can operate the two GENERAL alarm and HIGH alarm relays on the
CONTACT OUTPUT version of Sentro 1.
Gas Sensor
www.trolex.com
Summary of Contents for TX6351
Page 1: ......
Page 2: ......
Page 6: ...6 P5536 1600 rev A 1 4 Dimensions www trolex com...
Page 30: ...30 P5536 1600 rev A 4 Setup and Calibration 4 1 Controls and Indicators www trolex com...
Page 31: ...TX6351 TX6352 User Manual P5536 1600 rev A 31 www trolex com...
Page 55: ......
Page 56: ......
Page 57: ......
Page 58: ......