TX6351 • TX6352 User Manual
P5536.1600 rev A 17
Checkpoint - Carbon Dioxide Sensors
Normal atmosphere contains carbon dioxide which will affect the sensor signal so
nitrogen gas must be used to accurately calibrate the zero value. Power should
be applied to the sensor and the nitrogen applied for at least 5 minutes to ensure
that the sensor has stabilised.
Checkpoint
Not suitable for use with the TX6352 4 to 20 mA two wire, line powered version
of Sentro 1.