3.4 Connections
3.4.1 0.4 to 2 V Analogue Output
Signal
A low impedance two-wire voltage output
signal requiring a separate power supply to
the sensor.
12 V dc power can usually be derived from
the monitoring instrument being used.
The connection configuration works well up
to 5-100 m distance between the sensor and
the monitoring equipment, dependent on
cabling and sensor current consumption.
Both the signal and the power supply to the
sensor are being carried in the common 0 V
conductor so at some point – influenced by
the length of the cable and the resistance of
the cable cores – the current flowing in the
0 V conductor will impose an unacceptable
voltage error onto the signal.
This effect can be reduced on long distance
connections by increasing the size of the
cable cores, or even better, by running a
separate 0 V conductor to power the sensor
enabling operating distances up to 1,000 m.
Alternatively, the sensor may be powered by
a suitable power source close by.