11. Status codes

Technical data

external temperature sensor TS 910

Temperature

Operating principle

NTC

Measuring range

-20 °C to +80 °C

Accuracy

0.2 °C (0…40 °C), otherwise, 0.8 °C

Screen resolution

0.1 °C (0…40 °C), otherwise, 0.2 °C

Permissible operating 
ambient conditions

Air temperature

-20 °C to +50 °C

Relative humidity

0 to 100 % RH

Configuration 
and dimensions

Dimensions

Cable length 6 m, sensor length 20 mm, sensor diameter 5 mm, weight 85 g

Type of protection

IP65

Connection

3.5 mm plug socket

In the event of a sensor error, the following codes may appear on one of the three LCD rows instead of the measured value

E 2C

.........Initialisation error at the sensor

E 27

.........Faulty calibration data

E 36

.........Channel disabled (e.g. channel configured in the LCD, 

but sensor logged off or removed)

E 50

.........Value to be displayed too large to fit on display

E 51

.........Value to be displayed too small to fit on display

E 52

.........Channel value in the saturation (upper limit)

E 53

.........Channel value in the saturation (lower limit)

E 54

.........Data error. Received data are not plausible

E 55

.........Measurement not possible of sensor not reachable

Summary of Contents for DL200P

Page 1: ...DL200P EN TRT BA DL200P WM 02 EN OPERATING MANUAL DATA LOGGER ...

Page 2: ...etected at variable selectable recording or request intervals then saved and transmitted to a connected PC The measuring device may only be used for this intended use while complying with specified technical data Any other use is considered misuse and contrary to the intended use The product must not be disposed of with household waste Dispose of this device in a manner according to the relevant l...

Page 3: ...for active measured value recording Display for inactive measured value recording Display for date Display for time Mode selection button Data logger with internal sensors and additional connections for four external measured value sensors 3 5 mm plug socket for external temperature sensor 1 T1ext 3 5 mm plug socket for external temperature sensor 2 T2ext BNC connector for external building moistu...

Page 4: ...ger to your PC via the USB cable provided in the scope of supply The measuring device is automatically detected by the operating system Alternatively you can configure your data logger via a LAN connection over your local network if the network function is enabled Further information about the network function is provided in chapter 6 2 2 Start the SmartGraph software The program automatically det...

Page 5: ...ly configured with one button operation by using the mode selection button on your data logger You can restrict the one button operation with the mode selection button from your software if necessary key lock It is not possible to operate your data logger with the mode selection button in this case 6 1 Switching on and off When power is being supplied the data logger cannot completely switch off b...

Page 6: ...nd not connected to an external power supply it automatically deactivates the network function after 12 minutes pass without communicating with the network In this case the network function must be restarted by an external power supply either manually or automatically b When using the data logger in LAN mode the power should thus be supplied by the USB power supply In the optionally available PoE ...

Page 7: ...ing because it is not possible to calculate these values in the software m The dew point temperature is its own channel group and thus can not be calculated from the measured values of other channel groups retrospectively during a later evaluation So when organising the memory select the necessary dew point temperature measuring channels for recording as needed The specifications for type duration...

Page 8: ...eresis can be set by which the measured value must return to the valid range to switch off the alarm Example display for an alarm result of the measuring channel shown in the first measured value row Alarm display Upper alarm limit value Valid value corridor good range Lower alarm limit value flashing static Logging mode Reading the measured value memory 8 Alarm function 8 3 Acoustic alarm If an a...

Page 9: ...e is not possible and it does not respond to requests from the Smart Graph3 software After memory has been completely deleted FOMA FLSH 100 is shown on the display Afterwards the display returns to normal operation Further detailed information about using the software is provided in the software manual which you can open from the help function of the Smart Graph software Display for deleting measu...

Page 10: ...12 15 30 min 1 3 6 12 24 h Data storage integrated Flash memory with 16 MB capacity approx 3 200 000 measured values Data recording up to 20 simultaneous measuring channels Configuration and dimensions LCD display W 90 x H 64 mm Housing Plastic Dimensions L 166 x W 32 x H 78 mm Weight approx 250 g Interfaces USB LAN Connection for external sensors 2 x 3 5 mm plug socket Connection for external ele...

Page 11: ...ight 85 g Type of protection IP65 Connection 3 5 mm plug socket In the event of a sensor error the following codes may appear on one of the three LCD rows instead of the measured value E 2C Initialisation error at the sensor E 27 Faulty calibration data E 36 Channel disabled e g channel configured in the LCD but sensor logged off or removed E 50 Value to be displayed too large to fit on display E ...

Page 12: ...Trotec GmbH Grebbener Straße 7 52525 Heinsberg 49 2452 962 400 49 2452 962 200 info trotec com www trotec com ...

Reviews: