background image

Data logger operating manual

8

EN

8.1.    Alarm configuration

An alarm function can be configured in the measuring device administration for
each of the measuring channels selected for display or recording. 

By defining an upper and lower alarm limit value, a value corridor is specified.
This is the so-called good range and when the range is exited, an alarm sounds. 

Note: The alarm function can only be represented on the display for the chan-
nel  groups  of  the  three  measuring  channels  selected  for  measured  value
 display, and only in the operating modes M2 and M3 ! 

However, saving alarm results of other selected measuring channels in logging
mode  is  independent  of  the  selected  display  of  the  measured  values.  That
means  that  the  alarm  results  of  the  measuring  channels  which  are  not  for
display but for data recording are also recorded. 

Even  in  the  operating  mode  M4  when  displaying  measured  values  is  deacti -
vated, all preselected alarm results are recorded!

8.2.    Alarm display

The  alarm  display  is  only  possible  for
measured values which are preselected
for displaying and which have been con-
figured  with  an  active  alarm  function.

If an alarm occurs at one of these mea-
suring  channels,  the  alarm  symbol
flashes beside the measured value dis-
play  for  this  channel  while  the  good
range is exited. 

If the measured value then reaches the

defined  value  corridor,  that  is,  the  good  range,  then  the  alarm  symbol  stops
flashing and is shown continually. 

This indicates that an alarm result occurred. If the good range is exited during
the course of a measurement, the alarm symbol flashes again.

Reading the measured value memory deletes the alarm display.

Additionally, a hysteresis can be set by which the measured value must return
to the valid range to switch off the alarm.

Example display for an alarm result of the
measuring channel shown in the first
measured value row.

Alarm display

Upper alarm 
limit value

Valid value
 corridor 
“good range”

Lower alarm
limit value

flashing

static

Logging mode

Reading the measured value memory

8. Alarm function

8.3.    Acoustic alarm

If an alarm goes off while the acoustic function is activated (see chapter 6.2.4.
Acoustic function), then and alarm tone sounds and only stops when the mea-
sured value returns to the good range.

8.4.    Using the alarm hysteresis

If you use the alarm function without alarm hysteresis, then an alarm sounds
and is recorded each time the preset limit values are exceeded. 

If your limit values have been selected very close to each other, then this means
that an alarm situation occurs very often. 

For example, if you set a room temperature of 24 °C as the upper alarm limit
and a room temperature of 10 °C as the lower alarm limit and the room temper-
ature continually deviates between 23.5 and 25 °C during the measuring period,
then there will be many individual alarm signals created and recorded.

To prevent this, you can define an alarm hysteresis. With this setting, you define
a value by which the measured value must have returned to the valid value cor-
ridor, the good range, in order to switch off the alarm. 

If an alarm hysteresis of 1 °C was set, the alarm in the previous example would
go  off  once  as  24  °C  was  exceeded  and  then  only  stop  as  the  temperature
drops below 23 °C again.

H

Y

S

T

E

R

E

S

IS

Alarm display without hysteresis

Alarm display with hysteresis

Logging mode

Logging mode

Upper alarm 
limit value

Valid value
 corridor 
“good range”

Lower alarm
limit value

flashing

static

Upper alarm 
limit value

Valid value
 corridor 
“good range”

Lower alarm
limit value

flashing

static

Summary of Contents for DL200P

Page 1: ...DL200P EN TRT BA DL200P WM 02 EN OPERATING MANUAL DATA LOGGER ...

Page 2: ...etected at variable selectable recording or request intervals then saved and transmitted to a connected PC The measuring device may only be used for this intended use while complying with specified technical data Any other use is considered misuse and contrary to the intended use The product must not be disposed of with household waste Dispose of this device in a manner according to the relevant l...

Page 3: ...for active measured value recording Display for inactive measured value recording Display for date Display for time Mode selection button Data logger with internal sensors and additional connections for four external measured value sensors 3 5 mm plug socket for external temperature sensor 1 T1ext 3 5 mm plug socket for external temperature sensor 2 T2ext BNC connector for external building moistu...

Page 4: ...ger to your PC via the USB cable provided in the scope of supply The measuring device is automatically detected by the operating system Alternatively you can configure your data logger via a LAN connection over your local network if the network function is enabled Further information about the network function is provided in chapter 6 2 2 Start the SmartGraph software The program automatically det...

Page 5: ...ly configured with one button operation by using the mode selection button on your data logger You can restrict the one button operation with the mode selection button from your software if necessary key lock It is not possible to operate your data logger with the mode selection button in this case 6 1 Switching on and off When power is being supplied the data logger cannot completely switch off b...

Page 6: ...nd not connected to an external power supply it automatically deactivates the network function after 12 minutes pass without communicating with the network In this case the network function must be restarted by an external power supply either manually or automatically b When using the data logger in LAN mode the power should thus be supplied by the USB power supply In the optionally available PoE ...

Page 7: ...ing because it is not possible to calculate these values in the software m The dew point temperature is its own channel group and thus can not be calculated from the measured values of other channel groups retrospectively during a later evaluation So when organising the memory select the necessary dew point temperature measuring channels for recording as needed The specifications for type duration...

Page 8: ...eresis can be set by which the measured value must return to the valid range to switch off the alarm Example display for an alarm result of the measuring channel shown in the first measured value row Alarm display Upper alarm limit value Valid value corridor good range Lower alarm limit value flashing static Logging mode Reading the measured value memory 8 Alarm function 8 3 Acoustic alarm If an a...

Page 9: ...e is not possible and it does not respond to requests from the Smart Graph3 software After memory has been completely deleted FOMA FLSH 100 is shown on the display Afterwards the display returns to normal operation Further detailed information about using the software is provided in the software manual which you can open from the help function of the Smart Graph software Display for deleting measu...

Page 10: ...12 15 30 min 1 3 6 12 24 h Data storage integrated Flash memory with 16 MB capacity approx 3 200 000 measured values Data recording up to 20 simultaneous measuring channels Configuration and dimensions LCD display W 90 x H 64 mm Housing Plastic Dimensions L 166 x W 32 x H 78 mm Weight approx 250 g Interfaces USB LAN Connection for external sensors 2 x 3 5 mm plug socket Connection for external ele...

Page 11: ...ight 85 g Type of protection IP65 Connection 3 5 mm plug socket In the event of a sensor error the following codes may appear on one of the three LCD rows instead of the measured value E 2C Initialisation error at the sensor E 27 Faulty calibration data E 36 Channel disabled e g channel configured in the LCD but sensor logged off or removed E 50 Value to be displayed too large to fit on display E ...

Page 12: ...Trotec GmbH Grebbener Straße 7 52525 Heinsberg 49 2452 962 400 49 2452 962 200 info trotec com www trotec com ...

Reviews: