background image

DL200X operating manual for external sensors

11

EN

9. Alarm function

9.1.    Alarm configuration

An alarm function can be configured in the measuring device administration for
each of the measuring channels selected for display or recording. 

By defining an upper and lower alarm limit value, a value corridor is specified.
This is the so-called good range and when the range is exited, an alarm sounds. 

Note:  The  alarm  function  can  only  be  represented  on  the  display  for  the
 channel  groups  of  the  three  measuring  channels  selected  for  measured
value display, and only in the operating modes M2 and M3 ! 

However, saving alarm results of other selected measuring channels in logging
mode  is  independent  of  the  selected  display  of  the  measured  values.  That
means  that  the  alarm  results  of  the  measuring  channels  which  are  not  for
display but for data recording are also recorded. 

Even in the operating mode M4 when displaying measured values is deactivated,
all preselected alarm results are recorded !

9.2.    Alarm display

The  alarm  display  is  only  possible  for
measured values which are preselected
for displaying and which have been con-
figured  with  an  active  alarm  function.

If an alarm occurs at one of these mea-
suring  channels,  the  alarm  symbol
flashes beside the measured value dis-
play  for  this  channel  while  the  good
range is exited. 

If the measured value then reaches the

defined  value  corridor,  that  is,  the  good  range,  then  the  alarm  symbol  stops
flashing and is shown continually. 

This indicates that an alarm result occurred. If the good range is exited during
the course of a measurement, the alarm symbol flashes again.

Reading the measured value memory deletes the alarm display.

Additionally, a hysteresis can be set by which the measured value must return
to the valid range to switch off the alarm.

Example display for an alarm result of the
measuring channel shown in the first
measured value row.

Alarm display

Upper alarm 
limit value

Valid value
 corridor 
“good range”

Lower alarm
limit value

flashing

static

Logging mode

Reading the measured value memory

9.3.    Acoustic alarm

If an alarm goes off while the acoustic function is activated (see chapter 6.2.4.
Acoustic  function),  then  and  alarm  tone  sounds  and  only  stops  when  the
 measured value returns to the good range.

9.4.    Using the alarm hysteresis

If you use the alarm function without alarm hysteresis, then an alarm sounds
and is recorded each time the preset limit values are exceeded. 

If your limit values have been selected very close to each other, then this means
that an alarm situation occurs very often. 

For example, if you set a room temperature of 24 °C as the upper alarm limit
and a room temperature of 10 °C as the lower alarm limit and the room temper-
ature continually deviates between 23.5 and 25 °C during the measuring period,
then there will be many individual alarm signals created and recorded.

To prevent this, you can define an alarm hysteresis. With this setting, you define
a  value  by  which  the  measured  value  must  have  returned  to  the  valid  value
 corridor, the good range, in order to switch off the alarm. 

If an alarm hysteresis of 1 °C was set, the alarm in the previous example would
go  off  once  as  24  °C  was  exceeded  and  then  only  stop  as  the  temperature
drops below 23 °C again.

H

Y

S

T

E

R

E

S

IS

Alarm display without hysteresis

Alarm display with hysteresis

Logging mode

Logging mode

Upper alarm 
limit value

Valid value
 corridor 
“good range”

Lower alarm
limit value

flashing

static

Upper alarm 
limit value

Valid value
 corridor 
“good range”

Lower alarm
limit value

flashing

static

Summary of Contents for DL200X

Page 1: ...DL200X EN TRT BA DL200X WM 03 EN OPERATING MANUAL DATA LOGGER FOR EXTERNAL SENSORS...

Page 2: ...ss of the results presented is excluded from any liability or guarantee Liability for damages which have been caused by utilising the presented measured results is strictly excluded 2 Intended use The...

Page 3: ...ording Display for inactive measured value recording Display for date Display for time Mode selection button Data logger for connecting to external sensors 3 Equipment 5 pin round connector M12 for co...

Page 4: ...left and the display returns back to the original mode without making changes This function is only available if a BUS sensor has already been registered M61 M81 An 0 M1 M2 M3 M4 M51 M52 M53 M6x M81 A...

Page 5: ...essary that UDP broadcasts are allowed through the network b Note Based on their concept UDP broadcasts do not operate through a router NAT but instead only within its own network If the data logger i...

Page 6: ...e BUS sensor to the 5 pin round connector of the data logger see chapter 3 The sensor is supplied with current via the plug By optionally using up to three Y distributors a maxi mum of four digital BU...

Page 7: ...el 2 has pins 2 4 6 8 and 10 see chapter 3 Both of the input channels support the sensor configurations listed in the following subchapters Example display Registration function M61 selection for regi...

Page 8: ...r the first input channel Alternatively navigate to the configuration mode M81 AN1 to configure the specification of a sensor for the second input channel The option for configuring the first analogue...

Page 9: ...1 Further detailed information about configuration of the processing channels in the software is provided in the software manual which you can open by us ing the help function of the SmartGraph softw...

Page 10: ...on 0 20 mA Current mA cur min max mid Pt100 temperature measurement in 3 cable operation Temperature C cur min max mid Pt100 temperature measurement in 3 cable operation Temperature F cur min max mid...

Page 11: ...display Additionally a hysteresis can be set by which the measured value must return to the valid range to switch off the alarm Example display for an alarm result of the measuring channel shown in th...

Page 12: ...itioning for mobile use For mobile measured value recording the data logger can be positioned at any site Observe the permissible ambient conditions for operation see technical data Due to its compact...

Page 13: ...C Accuracy 0 2 C 0 1 of the measured value Resolution 0 02 C Memory organisation Sensing interval 10 30 s 1 10 12 15 30 min 1 3 6 12 24 h Saving interval 1 10 12 15 30 min 1 3 6 12 24 h Data storage 1...

Page 14: ...gative polarity approx 0 024 V 0x53 CHANNEL_UNDERRANGE Limit range 0 024 V 0 000 V Displayed 0 000V Normal operation mode 0 000 V 1 000 V Limit range 1 000 V 1 024 V Displayed 1 000V Overvoltage appro...

Page 15: ...Trotec GmbH Grebbener Stra e 7 52525 Heinsberg 49 2452 962 400 49 2452 962 200 info trotec com www trotec com...

Reviews: