Model 2701-B
3–5
To enter a new density offset, press
〈
3
〉
. The gauge prompts
for the density offset as shown below:
To change the offset, use the arrow keys to select the offset
sign (positive or negative). The gauge displays:
Enter the density offset as determined from the procedure on
page 3–3 and press
〈
ENTER/START
〉
. The gauge enables the
new offset and returns to the
Ready
screen.
Dens i t y Of f set
#. ## pcf
Sel ect ( +/ - )
Dens i t y Of f set
+ #. ## pcf
I nput and
Pr ess <ENTER>