Model 2701-B
4–3
4.
Press the
〈
ENTER/START
〉
key to continue. The gauge
stores the result of the reference and returns to the
Ready
screen. The result is then used to adjust all gauge
readings.
NOTE
If the reference result is out of range and the gauge
displays an error message, check the following:
Ensure that the circular sensor cover is firmly in place on
the gauge sensor (press firmly around the perimeter of
the cover).
Ensure that the cover and reference standard are totally
clean and free of any asphalt residue, moisture, or debris.
Clean with WD-40 or other suitable means.
Ensure that the sensor cover is seated flat and flush on
the reference standard, within the four locating pins, with
the gauge in the case.
Repeat the referencing process. If the reference result still
does not pass, contact Troxler technical support (see the
front of this manual for a list of service centers and their
contact information).