After turning the gauge on for the first time, set up the software
parameters described in the following sections. Several of these
parameters accessed from the gauge’s
Setup
and
Extended
menus. Other parameters are accessed via dedicated function keys.
The gauge stores the software setup, so the operator does not need
to enter a new setup each time the gauge is turned on. The setup
may be changed at any time, however.
The gauge offers a
Status
function that enables the operator to view
selected information concerning the current gauge status and setup.
To access the
Status
function, press the
〈
STATUS
〉
key. The gauge
displays two screens of information, including the measurement
units, count time, measurement mode, battery status, Gmb
(Marshall) value, Proctor value, Gmm (Voidless density) value, and
measurement depth. Use the arrow keys to scroll between the two
screens.
SETUP MENU
The gauge provides three different count times to be used in taking
measurements. The gauge can also display measurement results in
either U.S. units (pcf) or metric (SI) units (kg/m
3
or g/cm
3
). The
Model 3440 Plus gauge offers two depth modes:
automatic
and
manual
. These settings are accessed from the gauge’s
Setup
menu.
To display this menu, press the
〈
SETUP
〉
key. For information on
all of the functions available from the
Setup
menu, see Chapter 5.
Count Time
The
count time
defines how long the gauge reads (15 seconds, 1
minute, or 4 minutes). Longer count times produce better
measurement precision. Troxler recommends a count time of
1 minute for most sample measurements.
3. G
ETTI
NG
S
TA
RTE
D
Summary of Contents for 3430 Plus
Page 3: ...Model 3430 Plus 3440 Plus iii ...
Page 7: ...Model 3430 Plus 3440 Plus vii NOTES ...
Page 10: ......
Page 14: ...Model 3430 Plus 3440 Plus xii NOTES ...
Page 16: ...xiv NOTES ...
Page 30: ...2 6 NOTES ...
Page 64: ...4 NOTES ...
Page 98: ...7 NOTES ...
Page 122: ...9 NOTES ...
Page 142: ...Appendix A 20 NOTES ...
Page 178: ...Appendix D 2 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 180: ...Appendix D 4 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 190: ...Appendix F 8 NOTES ...
Page 194: ...Appendix G 4 NOTES ...
Page 200: ...Appendix H 6 NOTES ...
Page 215: ...Model 3430 Plus 3440 Plus Index 11 W Wide Area Augmentation System WAAS 2 Y YES key 3 4 INDEX ...
Page 216: ...Index 12 NOTES ...
Page 217: ...Model 3430 Plus 3440 Plus Warranty 1 NOTES WARRANTY ...