Model 3430 Plus & 3440 Plus
7–7
As prompted, place the gauge on the measurement site and press
any key. The gauge takes a 240-second count, during which it
displays the progress of the count. Upon completion, the gauge
calculates and displays the moisture offset (
K
) as shown:
To enable and use this moisture offset without storing it for later
use, press
〈
NO
〉
. The gauge enables the offset and returns to the
Ready
screen.
To enable the moisture offset and store it for later use, press
〈
YES
〉
. The gauge displays:
The gauge can store the value in one of four
memory cells
.
Storing a new value in a cell erases the old value. Use the
number keys to store the value. This will replace any stored
values that were previously associated with the memory cell.
The gauge stores the moisture offset, enables it for use, and
returns to the
Ready
screen.
DISABLE THE MOISTURE OFFSET
To disable the moisture offset value, press
〈
6
〉
at the
Moisture
Offset
menu shown on page 7–4. The gauge disables the moisture
offset and returns to the
Ready
screen.
NOTE
The mositure offset is also disabled when the gauge is
turned off.
7. C
AL
IBR
AT
IO
N O
FF
SET
S
Summary of Contents for 3430 Plus
Page 3: ...Model 3430 Plus 3440 Plus iii ...
Page 7: ...Model 3430 Plus 3440 Plus vii NOTES ...
Page 10: ......
Page 14: ...Model 3430 Plus 3440 Plus xii NOTES ...
Page 16: ...xiv NOTES ...
Page 30: ...2 6 NOTES ...
Page 64: ...4 NOTES ...
Page 98: ...7 NOTES ...
Page 122: ...9 NOTES ...
Page 142: ...Appendix A 20 NOTES ...
Page 178: ...Appendix D 2 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 180: ...Appendix D 4 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 190: ...Appendix F 8 NOTES ...
Page 194: ...Appendix G 4 NOTES ...
Page 200: ...Appendix H 6 NOTES ...
Page 215: ...Model 3430 Plus 3440 Plus Index 11 W Wide Area Augmentation System WAAS 2 Y YES key 3 4 INDEX ...
Page 216: ...Index 12 NOTES ...
Page 217: ...Model 3430 Plus 3440 Plus Warranty 1 NOTES WARRANTY ...