Model 3430 Plus & 3440 Plus
3–5
SOURCE ROD POSITIONS
As shown in Figure 3–3, the source rod can be placed in the
SAFE
,
backscatter, or direct transmission positions. When not taking
measurements, keep the source rod in the
SAFE
position. When
measuring thin layer or other materials through which you cannot
drill a hole, use the backscatter position. In the direct transmission
positions, the source rod extends into a pre-drilled hole.
WARNING
The source rod should always be in the SAFE
position when the gauge is not in use.
BS
2"
50mm
3"
75mm
4"
100mm
5"
125mm
6"
150mm
7"
175mm
8"
200mm
9"
225mm
10"
250mm
11"
275mm
12"
300mm
Figure 3–3. Source Rod Positions
(Maximum Depth of 300 mm in Increments of 50 mm)
3. G
ETTI
NG
S
TA
RTE
D
Summary of Contents for 3430 Plus
Page 3: ...Model 3430 Plus 3440 Plus iii ...
Page 7: ...Model 3430 Plus 3440 Plus vii NOTES ...
Page 10: ......
Page 14: ...Model 3430 Plus 3440 Plus xii NOTES ...
Page 16: ...xiv NOTES ...
Page 30: ...2 6 NOTES ...
Page 64: ...4 NOTES ...
Page 98: ...7 NOTES ...
Page 122: ...9 NOTES ...
Page 142: ...Appendix A 20 NOTES ...
Page 178: ...Appendix D 2 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 180: ...Appendix D 4 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Page 190: ...Appendix F 8 NOTES ...
Page 194: ...Appendix G 4 NOTES ...
Page 200: ...Appendix H 6 NOTES ...
Page 215: ...Model 3430 Plus 3440 Plus Index 11 W Wide Area Augmentation System WAAS 2 Y YES key 3 4 INDEX ...
Page 216: ...Index 12 NOTES ...
Page 217: ...Model 3430 Plus 3440 Plus Warranty 1 NOTES WARRANTY ...