Appendix A–2
TROUBLESHOOTING
WARNING
To prevent personal injury or equipment damage,
become familiar with the safety warnings on page
1–1 before operating the compactor.
Gyratory compactor does not turn on
1
1
.
.
Ensure that the unit is plugged in.
2
2
.
.
Check that the power switch is in the
ON
position.
3
3
.
.
Ensure the Emergency Stop button is not engaged.
4
4
.
.
Check the power to the wall outlet.
5
5
.
.
Check that the fuse for the wall outlet is not blown or that the
circuit breaker is not tripped.
Moving parts are not moving
1
1
.
.
Ensure the unit is turned on.
2
2
.
.
Check that the chamber door is closed.
3
3
.
.
A switch may be broken. Check the service panels and
emergency switches.
4
4
.
.
Verify that the hydraulic oil in the reservoir is at the proper
level, as described on page 1–12. Fill, if necessary, with light
grade (ISO VG 32, SAE 10) hydraulic oil.
Gyration data does not download to the computer
1
1
.
.
Check that the computer is properly connected to the
compactor’s serial port.
2
2
.
.
Check that the
Auto Output
feature is enabled (see page 3–11).
3
3
.
.
Check that the serial port is specified as the output device (see
page 3–12).
Summary of Contents for 5850v2
Page 10: ...x LIST OF TABLES Table Title Page Table 2 1 Control Panel Keys and Button 2 2...
Page 17: ...Model 5850v2 1 7 1 INTRODUCTION Figure 1 1 Model 5850v2 Gyratory Compactor...
Page 27: ...Model 5850v2 2 3 2 SETUP OPERATION Figure 2 1 Model 5850v2 Control Panel Layout...
Page 35: ...Model 5850v2 2 11 2 SETUP OPERATION part numbers 106952 or 106953...
Page 37: ...Model 5850v2 2 13 2 SETUP OPERATION the chamber door...
Page 62: ...4 8 Figure 4 3 Routing Load Cell Cables ROUTE CABLES THROUGH NOTCH LOAD CELL...
Page 68: ...4 14 NOTES...
Page 76: ...5 8 NOTES...
Page 86: ...Appendix A 10 NOTES...
Page 99: ...Model 5850v2 Index...
Page 100: ...NOTES...
Page 102: ...12...