SAFETY PRECAUTIONS

Only qualified person should operate, install and assure the maintenance. 
All International and European recommendations on electrical installation should be applied. 

•  The device may only be connected to supply power, which complies with the specifications included in the technical data. 
•  The device must be disconnected from all sources of power during installation and maintenance work. 
•  The device may only be operated under the conditions specified in the operating instructions.

FUNCTIONS - DESCRIPTION

The turbidity-measuring device determines the turbidity of liquids using the light absorption and scattered light measuring method.
The inline measuring cell at the sensor set makes use of two IR transmitters and one IR receiver in pulsating light mode (IR mea-
suring method according ISO 7027). The interconnected BAMOPHOX TUR measuring amplifier (option) analyses the 4 to 20 mA
signal generated by the sensor set. 

MOUNTING

- The fixture must be installed 

vertically

. – Best installation is the mounting in an ascending pipe with slow down section 600 mm

in front and 400 mm behind the turbidity-measuring device. 

- The sensor fixture must be completely filled with liquid during the performance of measurements. 

Install to the vertical portion of a siphon trap if necessary.

- Air bubbles distort measurement results. 
- Glass windows must be clean – clean them at regular intervals as required. 

SET UP – DISPLAYS

The sensor set consists of a receiver module with microprocessor-controlled analysis electronics and two emitter modules. 
The sensor set generates a 4 to 20 mA output signal in accordance with the following diagram. Measuring ranges are selected with
the help of the DIP switches on the receiver module. 

Page 2

446 M1 01 G

12-03-2015

Note: 

The tubidity measuring instruments are calibred with the internationally

specified standard suspension Formazin.

The indication takes places thus not in form of the measured light intensity, but
as concentration of the Formazin calibration suspension.

During  the  measurement  with  any  liquid,  this  means  thus  that  the  liquid
concerned causes the same light scattering as the standard suspension of
the indicated concentration.

Measured  values  of  other  turbidity  measuring  instruments,  wich  use  other
calibration  suspensions  and  measuring  angle, cannot  be  compared  direcly
with those with Formazin calibrated measuring instruments.

 

 

Signal 4-20 mA

FNU

Turbidité hors échelle

T

 

 

   

 

     

   

     

 

DIP-switch
range Set-up

Terminals

Calibration
potentiometers

Receiver module PCB

Over range

In  case  of  measure  out  of
the rang, the 4-20mA signal
is froze to 22mA

Indication

Green LED = 1Hz blinking

Ready, measuring in operation

Green LED = permanent ON

Ready, measuring not in operation

Green LED = OFF

No power supply, or defective

Fault indication

Red LED = ON

Measuring circuit faulty, (short circuit, or broken)
(court circuit ou rupture)

*) DIP5 and 6 

= not connected / no function

**) Armatures with nominal diameter > DN 65

Measuring range

DIP1

DIP2

DIP3

DIP4

DIP5* DIP6*

MB1 = 0.1... 50 FNU

OFF

OFF

OFF

ON

NC

NC

MB2 = 0.1...100 FNU

ON

OFF

OFF

ON

NC

NC

MB3 = 0.1...200 FNU

OFF

ON

OFF

ON

NC

NC

MB4 = 0.1...500 FNU (**300)

ON

ON

OFF

ON

NC

NC

MB5 = 100...1000 FNU

OFF

OFF

OFF

OFF

NC

NC