46
Messages
Display
Reason
Action
Invalid
Hematocrit
Repeat with new test strip,
using capillary whole
blood from the finger or
forearm. If error persists,
call 1-800-803-6025.
Temperature
Error
•
Too Cold/
Too Hot
Move meter and test
strips to area between
41°F-104°F; wait
10 minutes for system to
reach room temperature
before testing.
Sample Not
Detected or
Using Wrong
Test Test Strip
Retest with new
test strip and larger
sample.
Used Test Strip,
Test Strip outside
of vial too long,
Sample on top of
Test Strip.
Repeat with new test strip.
Make sure Sample Tip of
test strip touched top of
sample drop. If error
persists, call
1-800-803-6025.
Meter
Error
Call 1-800-803-6025