21
Guidance and manufacturer’s declaration –
electromagnetic
immunity
The model TS-6000 is intended for use in the electromagnetic
environment specified below. The customer or the user of the
model TS-6000 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment
–
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV
contact
±8 kV air
±6 kV
contact
±8 kV air
Floors should be
wood, concrete or
ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power
supply lines
±1 kV for
input/output
lines
±2 kV for
power
supply lines
±1 kV for
input/output
lines
Mains power
quality should be
that of a typical
commercial or
hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
Mains power
quality should be
that of a typical
commercial or
hospital
environment.
interruptions
and
voltage
variations
<5 % UT
(>95 % dip in
UT)
for 0,5 cycle
<5 % UT
(>95 % dip in
UT)
for 0,5 cycle