Important safety information
This section contains important safety information
regarding the lithium power cell. For more safety
information, refer to the safety manual and the user guide
supplied with the hearing instruments.
NOTICE
X
Do not exchange power cells. This may only be
performed by Service Center only.
Defective power cells
In case you think the power cell is defective, comply with
the following steps:
X
Place the hearing instrument into the charger. If you
get a red LED, then the power cell is defective.
X
Perform visual inspection of the hearing instrument.
The test is considered failed if it has any of below
criteria:
●
There is physical sign of deformation caused by
bloated power cell inside.
●
There is visible leakage from the power cell.
●
Product is generating heat on its own or hot to the
touch.