AutoVap15 Operating Manual | | v. 1.2
10 / 23
4.
Using the Vacuum Needle Manifold Valve Adjustment, adjust the vacuum pressure to
between 21-24 Hg.
****Note:Screwing in the needle valve (clockwise) will pull a deeper vacuum and unscrewing
(counterclockwise) will let more air into the system pulling less vacuum.
5.
Double Check the following:
Water Heater is set between 130-190F
Chiller is set between 40-44F
Now the AutoVap is ready to start processing
****Note: the vacuum gauge reads 30in Hg to 15 PSI.
4. Using the Black needle valve on the vacuum manifold,
adjust
vacuum pressure bewtween
21-24in Hg.
****Note:Screwing in the needle valve (clockwise) will pull a deeper vacuum and unscrewing
(counterclockwise) will let more air into the system pulling less vacuum.
STEP 2: PRIME FEED PUMP
Before turning your feed pump on, it is important to first prime the feed pump by allowing the vacuum to
"pull" your tincture through the pump and up through the preheater.