TruSteel AutoVap30 User Manual | | v. 1.9
18 / 39
VAC PSI = Vacuum Pump
If the vacuum pressure on the system "drops" to 10 PSI, the system will stop (lockout)
EVAP INLET = Water Heater
If the evaporator column senses the water heater temperature below 175°F (~79°C), the
system will stop (lockout)
This most likely means there is an issue with the water heater and it needs to be
addressed.
COND INLET = Chiller
If the condenser column senses the temperature above 55°F (~12°C), the system will
stop (lockout)
This most likely means there is an issue with the chiller and it needs to be addressed.
If one of these Lockouts are red, please verify whether the corresponding piece of equipment is
operating properly.