13
OPERATING MANUAL
RESISTANCE MEASUREMENT
WARNING
!
• To avoid damaging the meter or to the device under test, disconnect
circuit power and discharge all the high-voltage capacitors before
measuring resistance.
• To avoid electric shock, do not input higher than DC 60V or AC 30V
voltages.
1. Switch the function selector to
Ω
position (range: 600Ω/6kΩ/60kΩ/600MΩ/60MΩ), make
sure the circuit power is turned off.
2. Connect the common test lead to COM terminal then insert the red test lead to
Hz
V
Ω
°
C
terminal.
3. Connect the leads to the correct test points in the circuit to measure voltage.
NOTE:
•
If the measured resistor is open or the resistance exceeds the maximum range, the “OL”
symbol will be displayed on the screen.
•
When measuring low resistance, the test leads will produce 0.1Ω-0.3Ω measurement error. To
obtain accurate measurement, short-circuit the test leads and use the REL function.
•
If the resistance is greater than 0.5Ω when the test leads are shorted, please check if the test
leads are loose or damaged.
•
When measuring high resistance at 60MΩ range, it takes a few seconds to stabilise the
readings.
•
The internal 600mA and 20A fuses can be checked by the 6MΩ measurement function.
Please refer to Picture 5b for more details: Insert the red probe to the 630mA or 20A input
terminal to measure the resistance. If both fuses are blown, the “OL” symbol will appear on
the screen.
R
True RMS
OFF
Live
NCV 600m
6
60
60
60m
60m
°C°F
Hz%
hFE
600
6k
60k
600k
100mF
6M
60M
60
µ
6
600
600
600m
600m
20
20
1000
1000
V
Ω
A
V
A
Hz
Hz
°C
mA µA
20A
MAX
FUSED
250V max
600mA max
MAX 10sec EACH 15min
FUSED
250V max
20A max
VΩ
Hz
COM
CAT II 1000V
CAT III 600V
SEL
REL
OFF
A
HOLD
NPN
PNP
E
c
b
e
B
C
True RMS
OFF
Live
NCV 600m
6
60
60
60m
60m
°C°F
Hz%
hFE
600
6k
60k
600k
100mF
6M
60M
60
µ
6
600
600
600m
600m
20
20
1000
1000
V
Ω
A
V
A
Hz
Hz
°C
mA µA
20A
MAX
FUSED
250V max
600mA max
MAX 10sec EACH 15min
FUSED
250V max
20A max
VΩ
Hz
COM
CAT II 1000V
CAT III 600V
SEL
REL
OFF
A
HOLD
NPN
PNP
E
c
b
e
B
C
TTIDM1000V
TTIDM1000V