1
、
switching the cooling mode
The camera's default cooling mode is air-cooled. You can switch into water-cooled
by adjusting the fan gear. Every time you select fan Off, you will be prompted as
follows: click OK to close the fan
2
、
Cooling water selection: bottled water is recommended.
3
、
Water temperature
:
The water temperature in circulator is recommended to be
20
℃
。
4
、
Install water-cooled pipe correctly to ensure no water seepage at the interfaces
between the camera end and the circulator.
5
、
Before the operation of the camera, ensure that the installation of circulator and
camera water valve is correct, and the water flow in the water pipe reaches 1L/min.
6
、
Cooling water circulator
:
Use the circulator and cooled water correctly according
to its instructions
3.3 Installation precautions
- 11 -
- 12 -
7
、
According to the experimental data, condensation water will occur in water valves
and water pipes under inappropriate environmental conditions, and there is a
potential risk of damage to the equipment. In order to ensure the normal operation of
the equipment, the following provisions are made on the ambient temperature and
humidity (refer to the figure below).
8
、
During the operation of the camera, do not turn off the cooling water circulator
and stop the refrigeration.
9
、
After the operation, turn off the power of the camera and the circulator, and drain
their water.
Warning
:
Water-cooled and fan can be turned on at the same time, but not turned off at the
same time. At least ensure it is in operation with one cooling mode. Otherwise, the
sensor may be damaged by continuous high temperature operation