![background image](http://html1.mh-extra.com/html/tum-electronic/mlb1006/mlb1006_user-manual_1179838017.webp)
MLB1006 Serisi
- 13 -
Ex. When display shows [p 0.000], Press “
” key, the division set as 0.000, then enter the
maximum capacity setting.
3.1.3. Maximum capacity setting:
Display shows [ FULL ] around 2-3 seconds then display shows [000.000] the decimal point was set
as 3 decimal points, then press “
TARE
” key to move the flashing digits to the right in circle to select
the digits that will increase the number, then press “
ZERO
” key to increase the number until the
number appears, then repeat the above moves until display shows the maximum capacity then
press “
” key to enter calibration setting.
Ex. Display shows [50.000] then press “
” key to enter calibration setting.
3.1.4. Zero point calibration:
Display shows [CaL 0 ] Do not put any objects on the pan when display shows this, then scale
will move to next calibration setting.
3.1.5. Second point calibration: (Unit is in kg/ lb)
Display shows [00.000] the decimal point was set as 3 digits. (This depends on what user sets
in the last moves). Put on the weights on the pan which will be calibrate as second point then use
“
TARE
” key to move the flashing digits to the right in circle to select the digits that will increase the
number, then press “
ZERO
” key to increase the number until the number appears, then repeat the
above moves until display shows the numbers which is equal to the weights put on the pan, then
wait until the scale is stable then press “
” key, scale will move into the weighing mode then
press “calibration” key to confirm calibration finished.
3.2. Fast calibration mode:
Press and hold “
” key, then turn on the scale, then press “
FUNCTION
” key to enter
calibration mode: (the division, decimal point and max. point will remain the same)