Any material to be tested should be clean and free from grease, finger marks 
and dust. If necessary the material should be cleaned using a propriety glass 
cleaner and thoroughly dried using lint free cloth.

Before use always check and adjust the Tintman instrument sensitivity against the 
sample of 

Reference Window

 supplied in the instrument kit. The 

Reference Window

 

is marked with the transmission measured by a reference instrument. Please refer to 
the 

Checking and Adjusting Tintman

 instructions at the end of this guide.

To use the instrument connect the light source and detector to the readout unit 
with the colour coded connectors. Check the colours match and the connectors 
are pushed in firmly. The light source has a red connector, the detector black. No 
damage will be done if the connectors are mistakenly swapped.

Attach the source and detector to either side of the material to be measured, if 
necessary use the captive suckers to secure to vertical windows. The source and 
detector should self align magnetically. On raked windscreens attach the detector 
to the inside of the glass hanging from its sucker. When the source is placed on 
the outside of the windscreen the magnetic attraction between the two should 
automatically draw the detector to the windscreen.

Before taking a reading, visually check that the source and detector are 
concentrically aligned and firmly against the material.

To take a transmission reading press the large button on the readout unit. The 
display will show the measured transmission in percent. Allow about 10 to 20 
seconds with the button depressed for the light source intensity to stabilise. Keep the 
button pressed throughout this period. It is recommended that three readings are 
taken at different positions and the average value used.