background image

6

Thorough Cleaning (refer to exploded view)

Thorough Cleaning (refer to exploded view)

Thorough Cleaning (refer to exploded view)

Thorough Cleaning (refer to exploded view)

Thorough Cleaning (refer to exploded view)
If meter is plugged due to hardened chemical or debris, do
the following:
1. Drain all fluid from meter.
2. Unscrew meter cap (item 1).  Use a  strap type oil filter

wrench or large 5" jaw pipe wrench if necessary.

3. Insert a flat-head screwdriver in the lower slot (see Figure 7)

and turn to pry up meter cover (item 6).

3. Remove old batteries and insert new batteries, making sure

battery polarity is correct, or meter damage could occur.

4.  As noted in Figure  5, reinstall o-ring on electronics module.

Align sensor receptacle in proper location. Press module
gently down into meter cover.

Figure 6

1/8" GAP

CAP

RIDGE ON
OUTLET PORT

5. Thread on meter cap until hand tight. To check tightness,

there should be 1/8" gap between cap and ridge on outlet
port. (See Figure 6).

O-Ring

Figure 5

REPAIR

If any meter components are damaged, they should be re-
placed.  See meter drawing and parts list for correct replace-
ment part number before ordering.

Meter should be flushed between uses with water to prevent
chemicals from drying and plugging meter.

MAINTENANCE

Follow fluid manufacturer's recommended procedures
for handling and disposing of metered fluids.

CAUTION

Figure 7

Lower Pry Slot

4. The meter chamber (item 8) can now be removed.
5. Rinse all meter components with flushing fluid.  Do NOT

submerge display assembly.  Be careful not to get any fluid
or dirt in the electronics module.

6. Reassemble meter.

CALIBRATION NOTE:  Over time, the chamber inside the 825
meter will wear, requiring the meter to be recalibrated with
water.  When this should be done depends on the
amount and type of fluid dispensed.  In most crop protection
fluid uses (less than 1000 gallons of a clean fluid per year), the
825 meter will remain accurate for many years without
recalibration.  On the other hand, dispensing an abrasive
fluid may require more frequent recalibration.

The 825 meter is designed to be calibrated with clean water
for  safe handling. See "Water Calibration" section in  Appen-
dix-B.

     NOTE

:  Low battery icon will flash when batteries begin to

lose power.  Meter still functions properly for several days after
the icon begins to flash.  Neither calibration, current total or
totalizer quantities will be lost when replacing batteries.

To Replace Batteries

To Replace Batteries

To Replace Batteries

To Replace Batteries

To Replace Batteries (refer to exploded view)
1. Unscrew meter cap (item 1).  Use a strap type oil filter wrench

or large 5" jaw pipe wrench if needed.

2. Insert a flat-head screw driver into the  top slot (see Figure

4) and gently pry up electronics module.

CAUTION: Be careful not to get fluid or dirt in electronics area.

Figure 4

Upper Pry Slot

 

      CHANGING THE CAL FACTOR

CHANGING THE CAL FACTOR

CHANGING THE CAL FACTOR

CHANGING THE CAL FACTOR

CHANGING THE CAL FACTOR

•  Hold button  3  until the

display only shows CAL and
number.

•  Press   3  repeatedly until

you reach the desired
number.  Note – number will
increment up to 19, then
back to zero.

•  Press   2   to return to

normal operating mode.

Storage

Storage

Storage

Storage

Storage
Store in a cool, dry place.  Drain out all fluid that could freeze
in the meter.

BATTERIES

Summary of Contents for 850

Page 1: ......

Page 2: ......

Page 3: ......

Page 4: ...e fluid the the fluid the the fluid the the fluid the lower lower lower lower lower the CAL number the CAL number the CAL number the CAL number the CAL number The The The The The thicker thicker thicker thicker thicker the fluid the the fluid the the fluid the the fluid the the fluid the higher higher higher higher higher the CAL number the CAL number the CAL number the CAL number the CAL number C...

Page 5: ......

Page 6: ...yfluid or dirt in the electronics module 6 Reassemblemeter CALIBRATIONNOTE Overtime thechamberinsidethe825 meter will wear requiring the meter to be recalibrated with water When this should be done depends on the amount andtypeoffluiddispensed Inmostcropprotection fluiduses lessthan1000gallonsofacleanfluidperyear the 825 meter will remain accurate for many years without recalibration On the other ...

Page 7: ...ure sometimes change See Check Meter in Appendix A Meter reads low by Wrong calibration factor Use a lower calibration factor Chemical formulations 10 or less See 1 Step procedure sometimes change Temperature also affects accuracy Meter reads low by Meter chamber is worn Recalibrate meter with water See more than 10 Chamber is partially Appendix B See Check Meter in plugged Appendix A Clean chambe...

Page 8: ......

Page 9: ... and 4 atthesametimeandholdfor3seconds Display will read the version of the software loaded in the meter example 7 r1 02 2 Press ON to display accumulated total If over 9999 display will scroll across the screen 3 Holdbuttons ON and 4 for5secondstoresetaccumu lated total 4 Press button 2 twice to get back to normal operating mode You can check the calibration in your meter 1 Set to CAL 4 See to ch...

Page 10: ...sing press the ON button The left digit of the display will blink 6 6 6 6 6 Press the 4 button to increment the digit to the amount of fluid dispensed example 05 00 Press the ON button to move to the right If you make a mistake press button 2 to start back at the left most digit 7 7 7 7 7 After number is loaded press the ON button again to accept 8 8 8 8 8 Display now shows Since you are calibrati...

Page 11: ...fter dispensing press the ON button The left digit of the display will blink 7 7 7 7 7 Press the 4 button to increment the digit to the amount of fluid dispensed example 05 00 Press the ON button to move to the right If you make a mistake press button 2 to start back at the left most digit 8 8 8 8 8 After number is loaded press the ON button again to accept 9 9 9 9 9 Display now shows This is the ...

Page 12: ...ISCOSITY VISCOSITY CPS Low Viscosity thin Fluids CALIBRATION FACTOR VISCOSITY CPS VISCOSITY CPS CALIBRATION FACTOR Medium Viscosity Fluids High Viscosity very thick Fluids NOTE Graphs are accurate with original factory calibration or a water calibration See page 8 APPENDIX D ...

Reviews: