Rebar Scanner A60+
9
right side of Signal; saved amount of information is showed on the
right side of Memory; estimated rebar diameter at the current probe is
showed on the right side of Diameter; equivalent covering layer
thickness at the current probe is showed on the right side of Thickness .
Length of black scroll bar indicates the trend which probe is close to
the top of rebar. When black scroll bar is lengthened, it indicates that
probe is close to the top of rebar. When black scroll bar is shortened, it
indicates that probe is away from the top of rebar.
Figure 3-6 interface of diameter test
To find the position on the top of rebar and detect accurately
detected rebar diameter and covering layer thickness,
firstly, roughly
scan. After hearing the alarm, shift the probe backwards. Because the
speed of shifting the probe for the first time is too much, it is possible
to miss the data during collecting the data. Therefore, after audible
alarm, during shifting the probe backwards, slow down as much as
possible. When audible alarm is heard for the second time, data on the
right side of Signal are changed. Repeat such operation till the value
on the right side of Signal is the maximum and black scroll bar is the
longest. At this time, center of diamond pattern on the probe is just on
the top of rebar. Then press the
RANGE
key and slightly wait for a
No.=
Signal
Memory
Diameter
Thickness