Rebar Scanner A60+
13
select the position in the middle of crossing point of grid pattern
horizontal rebar to avoid the influence on detected longitudinal rebar
by grid pattern horizontal rebar. The trolley held by hand is moved
horizontally from the left to the right (the trolley is vertical to
extension direction of longitudinal rebar and forward speed is not
more than 20mm/s). One black block shown on the screen is moved
horizontally from the left to the right. After hearing the alarm, it
indicates there is rebar under the probe and rebar is showed on the
screen in the formation of being vertical to X-axis. At the same time,
covering layer thickness is showed under the T. During continuing to
move the trolley forwards, when horizontal length that the trolley
moves is not less than 500, it is prompted by means of continuous
alarm. At this time, press the↓ key to change the direction into↓. Then
detect the horizontal rebar of grid pattern. Similarly, detected point
should select the position in the middle of crossing point of grid
pattern longitudinal rebar to avoid the influence on detected horizontal
rebar by grid pattern longitudinal rebar. The trolley held by hand is
moved from up to down (pay attention to the direction of trolley). One
black block shown on the screen is also moved from up to down. After
hearing the alarm, detected rebar is showed on the screen in the
formation of being parallel to X-axis. At the same time, covering layer
thickness is showed under the T. During continuing to move the
trolley downwards, when longitudinal length that the trolley moves is
not less than 500, it is prompted by means of continuous alarm. At this
time, press the
SAVE
key to save the data.