61
Accuracy Table
Range
Measurement
range
AC voltage range
6.000 V
60.00 V
600.0 V
9.999 kHz 1.000 kHz to
9.999 kHz
0.600 V or
more
6.00 V or more 60.0 V or more
• The input is up to 3 × 10
6
V • Hz.
• “
−−−−
” appears when no measurement can be made.
*1: The measurement range from 5.00 Hz is only for the 6.000 V range.
The measurement range for other voltage ranges is 40.00 Hz to 99.99 Hz.
3
DC voltage
Range
Accuracy
Input impedance
600.0 mV
±0.5% rdg. ±5 dgt.
11.2 M
Ω
±2.0%
6.000 V
±0.5% rdg. ±5 dgt.
11.2 M
Ω
±2.0%
60.00 V
±0.5% rdg. ±5 dgt.
10.3 M
Ω
±2.0%
600.0 V
±0.5% rdg. ±5 dgt.
10.2 M
Ω
±1.5%
• Overload protection: 660 V DC/660 V AC or 3 × 10
6
V • Hz (energized
for 1 minute) (DT4221, DT4222)
• Overload protection: 750 V DC/750 V AC or 3 × 10
6
V • Hz (energized
for 1 minute) (DT4223, DT4224)
• Auto range movement threshold: 6000 counts or more for upper range,
540 counts or less for lower range.
4
AUTO V
Range
Accuracy*
1
Input impedance
DC, 40 to 500 Hz
Over 500 Hz to 1 kHz
600.0 V ±2.0% rdg. ±3 dgt. ±4.0% rdg. ±3 dgt.
900 k
Ω
±20%
• Overload protection: 660 V DC/660 V AC or 3 × 10
6
V • Hz (energized
for 1 minute) (DT4221, DT4222)
• Overload protection: 750 V DC/750 V AC or 3 × 10
6
V • Hz (energized
for 1 minute) (DT4223, DT4224)