62
Accuracy Table
• Transient overvoltage: 6000 V
• Crest factor: The crest factor is 3 up to 4000 counts and reduces linearly to
2 at 6000 counts.
• Connection method: DC coupling
*1: For AC voltage, the accuracy is speci
fi
ed in 1% or more of the range,
however, ±5 dgt. should be added to 5% or less of the range.
• Accuracy guarantee range for frequency: 40 Hz to 1 kHz (Measured
values outside the accuracy guarantee range for frequency are also
displayed.)
The accuracy is not speci
fi
ed for strain waveforms outside the range of
40 Hz to 1 kHz.
• For 100 Hz with the
fi
lter ON, ±1.5% rdg. is added to the accuracy
speci
fi
cation between 40 Hz and 100 Hz and the accuracy is not
speci
fi
ed in 100 Hz or more.
• For 500 Hz with the
fi
lter ON, ±0.5% rdg. is added to the accuracy
speci
fi
cation between 40 Hz and 500 Hz and the accuracy is not
speci
fi
ed in 500 Hz or more.
5
Continuity
Range
Accuracy
Measurement current
600.0
Ω
±1.0% rdg. ±5 dgt.
200
μ
A ±20%
• Open circuit voltage: 1.8 V DC or less (DT4221, DT4222)
2.0 V DC or less (DT4223, DT4224)
• Overload protection: 600 V DC/600 V AC or 3 × 10
6
V • Hz (energized for 1 minute)
Current under overload: Steady state 15 mA or less, transient state 0.8 A or less
(DT4221, DT4222)
• Overload protection: 750 V DC/750 V AC or 3 × 10
6
V • Hz (energized for 1 minute)
Current under overload: 1.5 mA or less
(DT4223, DT4224)
• Continuity ON threshold: 25
Ω
±10
Ω
(continuous buzzer sound)
• Continuity OFF threshold: 245
Ω
±10
Ω
• Accuracy guarantee condition: After zero adjustment has been performed