Display
Description
Action
Test result is HIGHER than 600 mg/dL.
Repeat the test by using a new test strip. If the same
message displays again, the result is confirmed to be higher
than the meter test range. Please consult your healthcare
professional immediately for further advice.
Test result is LOWER than 20 mg/dL.
Repeat the test by using a new test strip. If the
same message displays again, the result is
confirmed to be lower than the meter test range.
Please consult your healthcare professional
immediately for further advice.
Temperature is TOO HIGH during the
test procedure.
Large variation may occur between results due to high or
low temperature. Move to an environment (50°F ~ 104°F or
10°C ~ 40°C) and wait for 30 minutes before re-testing.
Temperature is TOO LOW during the
test procedure.
Large variation may occur between results due to high or
low temperature. Move to an environment (50°F ~ 104°F or
10°C ~ 40°C) and wait for 30 minutes before re-testing.
34