4. Press and hold for 2 seconds to change to the Control Solution test mode. " " will appear on the display
(Fig.43).
(with the voice function on, you will hear "Please apple control solution.")
Press again for 2 seconds and
the meter will switch back to the normal testing mode.
(with the voice function on, you will hear " Please apply blood
sample.")
5. Discard the first drop of the Control Solution. Squeeze a small drop on a clean nonabsorbent surface e.g. wax paper
Do not apply the Control Solution to the test strip directly from the bottle because contamination may occur (Fig.44)
6. Introduce the tip of the test strip to the Control Solution drop. The Control Solution will be automatically drawn into
the strip. Hold until the meter beeps. The meter will now start to count down and the Control Solution test result will
appear on the LCD screen (Fig.45).
(with the voice function on, you will hear "Control solution test result is ... milligram
per deciliter.")
(Fig.45)
(Fig.44)
(Fig.43)
Caution:
Follow the above instructions when performing a
Control Solution test in order to prevent contamination.
The Tyson Bio HT100-C Blood Glucose Monitoring System is working properly only if the Control Solution test result is
within the specified range printed on the test strip vial. Please repeat the test when the test result is out of range.
Out of range result may be caused by:
˙
Incorrect steps while performing the test
˙
The Control Solution temperature is lower than 59°F (15°C) or higher than 95°F (35°C)
˙
Expired or contaminated Control Solution
˙
Expired or contaminated test strips
˙
Meter malfunction
Control Solution Results
18