background image

PAGE 5 OF 18 

0623 IH-7935

OPERATION

POWER ON/OFF

1.   Press  the 

ON/OFF

 key for two seconds to turn the 

scale on or off. 

CAUTION! Never power the scale on with weight 
on the platform.  

2.  When turning on, the indicator will show  

000000-999999 before entering weighing mode. 

3.  Check that the LCD display and the status lights are 

working properly.

ZERO

INITIAL ZERO SETTING

When turning the scale on, if the weight on the display 

indicator is within the initial zero tolerance, the indicator 

will show 

zero

.

MANUAL ZERO SETTING

1.  Wait for the scale to stabilize.

2.  Press the 

ZERO

 key. The display will show zero weight.

TARE

1.  Press the 

TARE

 key. 

2.  The indicator will show the net weight. The Net and 

Tare indicators will be displayed on the bottom of 

the indicator. 

3.  Press the 

TARE

 key. Zero the weight. 

4.  The indicator will display the gross weight.

ACCUMULATION 

1.  Zero the weight. Load weight until the scale stabilizes.

2.  Press the 

TOTAL

 key to enter accumulation mode. 

3.  Total indicator will be displayed on the bottom of the 

indicator. Display will read 

n 001

 then will display 

the loaded weight. 

4.  Unload the weight. Zero the weight. Load the 

second weight until the scale stabilizes. 

5.  Press the 

TOTAL

 key. The display will read 

n 002

then display the second loaded weight. 

6.  Repeat a maximum of 999 times.

CHECK THE ACCUMULATION

1.  Press and hold the 

PRINT

 key. Press the 

TOTAL

 key.

2.  Display will read 

n**

, then will show total weight.

  NOTE: There are 8 digits total. The display 

shows the first 4 digits, then the last 4 digits.  
For example, when the first 4 digits displayed 
are "0012" and the last 4 digits displayed are   

 "34.56," the weight is "1234.56."

EXIT ACCUMULATION

1.  When the indicator shows the last 4 digits, press and 

hold the 

TOTAL

 key. The indicator will read 

clr n

2.  Press the 

PRINT

 key to exit. If you want to clear 

the total weight, press the 

ZERO

 or 

TARE

 key. The 

indicator will read 

clr y

.

3.  Press the 

PRINT

 key to clear the total weight and exit 

accumulating mode.

PRINT

1.  If the weight is stable, connect the scale to the 

printer. 

2.  Press the 

PRINT

 key. 

NOTE: While in tare mode, print with tare. If the 
scale shows a negative weight, printing is not 
allowed. 

COUNTING

Use this mode to count parts of uniform weight.

1.  In weighing mode, remove all weight from the 

platform. 

2.  Press the 

COUNT

 key to enter the counting mode.  

PCS 0

 will be shown on the display. 

3.   Press  the 

ZERO

  key  to  change  the  sample  quantity        

 – 5, 10, 20, 50, 100, 200 or 500 units. 

4.  Place the sample quantity of the item being 

counted on the scale platform and press the 

PRINT

 

key. Sample quantity will be shown on the display.

5.  Place remaining objects on the platform. The total 

quantity will be shown on the display.

6.  Press the 

COUNT

 key to return to weighing mode.

7.  To weigh different weights, at weighing mode, put 

the item on the scale and press the 

COUNT

 key. The 

indicator will read 

0

8.  Press and hold the 

COUNT

 key. Press the 

ON/OFF

 key. 

The indicator will show 

PCS 0

9.  Press the 

ZERO

 key and input the sample quantity. 

10.  Press the 

PRINT

 key to enter. Repeat steps 2 and 3.

Summary of Contents for H-10903

Page 1: ...against possible damage Running the instrument cable through conduit is the best method of protection UNPACKING 1 Inspect your shipment for damage If you see visible signs of damage notify carrier at...

Page 2: ...isplay Indicator 1 H 754 7510CB 65088 5 Load Cell 5 000 lb Capacity 1 H 5261 307301 5 Load Cell 10 000 lb Capacity 1 H 5262 307501 6 Load Cell Bolt 2 51601 7 Foot 4 H 4602 SFEET 50431 8 Access Plate 1...

Page 3: ...f the display the battery needs to be recharged Plug the power plug into the wall to charge the battery Allow 3 4 hours to fully charge battery Scale does not need to be turned on BATTERY MAINTENANCE...

Page 4: ...k to net weight automatically Tare At gross weight mode get the tare weight At net weight mode clear the tare get the gross weight Zero Zeros the weight within tolerance Print Works with ZERO TARE ON...

Page 5: ...weight NOTE There are 8 digits total The display shows the first 4 digits then the last 4 digits For example when the first 4 digits displayed are 0012 and the last 4 digits displayed are 34 56 the w...

Page 6: ...input and output nnnnnnn Calibration error Bad connection Load cell has quality problem Check that scale is level Check load cell connection Inspect load cell Check input and output Maintain a good co...

Page 7: ...instrumento a trav s de un conducto es el mejor m todo de protecci n DESEMPACAR 1 Inspeccione el cargamento en busca de da os Si observa se ales de da os visibles notifique al transportista de inmedi...

Page 8: ...e Pantalla LP 7510 1 H 754 7510CB 65088 5 Celda de Carga Capacidad de 5 000 lb 1 H 5261 307301 5 Celda de Carga Capacidad de 10 000 lb 1 H 5262 307501 6 Perno de la Celda de Carga 2 51601 7 Pata 4 H 4...

Page 9: ...cha de la pantalla la bater a deber cargarse de nuevo Enchufe a la pared para cargar la bater a Permita de 3 4 horas para cargar la bater a completamente No necesita encender la b scula MANTENIMIENTO...

Page 10: ...apagar Acumulaci n Funciona con el bot n PRINT para activar la funci n de acumulaci n y verificar el resultado acumulado Conteo Operaci n de conteo Conversi n a Lbs Kgs Convierte entre lbs y kgs Peso...

Page 11: ...itos en total La pantalla muestra los primeros 4 d gitos luego los ltimos 4 Por ejemplo cuando los primeros 4 d gitos que se muestren sean 0012 y los ltimos 4 d gitos mostrados sean 34 56 el peso es d...

Page 12: ...y entradas nnnnnnn Error de calibraci n Mala conexi n La celda de carga tiene un problema de calidad Verique que la b scula est nivelada Verifique la conexi n de la celda de carga Inspeccione la celda...

Page 13: ...re passer le c ble par un conduit est le meilleur moyen de le prot ger D BALLAGE 1 Inspectez l emballage pour rep rer toute trace de dommage Si vous remarquez des signes visibles d endommagement infor...

Page 14: ...10 1 H 754 7510CB 65088 5 Cellule de charge Capacit de 5 000 lb 1 H 5261 307301 5 Cellule de charge Capacit de 10 000 lb 1 H 5262 307501 6 Boulon de cellule de charge 2 51601 7 Pied 4 H 4602 SFEET 504...

Page 15: ...appara t sur l affichage en bas droite la batterie doit tre recharg e Branchez la fiche d alimentation au mur pour le rechargement de la batterie Pr voyez 3 4 heures pour que la batterie soit compl te...

Page 16: ...Net Affichage ACL Touches TOUCHE NOM DE LA TOUCHE FONCTION DE LA TOUCHE Mise en marche arr t Appuyez deux secondes pour mettre en marche arr t Accumulation Fonctionne avec la touche PRINT pour remplir...

Page 17: ...tal L cran affiche les 4 premiers chiffres puis les 4 derniers Par exemple lorsque les 4 premiers chiffres affich s sont 0012 et les 4 derniers 34 56 le poids est de 1234 56 QUITTER LA FONCTION ACCUMU...

Page 18: ...la sortie nnnnnnn Erreur d talonnage Mauvaise connexion Probl me de qualit de la cellule de charge V rifiez que la balance est de niveau V rifiez la connexion de la cellule de charge Inspectez la cel...