PAGE 3 OF 3
0816 IH-1101
ERROR MESSAGES
CAUSES
RECOMMENDATIONS
-- -- -- Blinking
Scale is in sleep mode
Press any key to change the
weight to resume normal
weighing. (Pressing a key
will exit the scale from
sleep mode and perform its
normal function.
-------- Not blinking
Scale waiting for weight to
stabilize
DISE
Display error
Scale is unable to display the
number completely
If no weight is on the scale,
press ZERO to return the
weight reading to zero.
LB
Low battery indicator
Blinks if left too long, scale shuts
itself off
Change or recharge
batteries.
OE
The scale is over capacity or
outside the A/D converter range
Reduce load on scale. If OE
continues to display, load
cell wiring or circuit board
chip may be bad.
EEPE
EEPROM error
Calibration is corrupted
or gone. Scale requires
recalibration.
TROUBLESHOOTING