PAGE 4 OF 12
0922 IH-1679-MNBRD
INSTRUCTIONS CONTINUED
CALIBRATION
11. Secure indicator faceplate to housing using four
previously removed screws and Phillips screwdriver.
(See Figures 19-20)
12. Complete calibration. (See Calibration below)
ENTER TECHNICAL PARAMETER MODE
NOTE: Must use 10% of capacity to calibrate.
1. Press
ZERO
and
TARE
keys at the same time. The
display will show "F1 UNIT".
2. Press
TARE
key until display shows "PROG" and press
ZERO
key to show "PN" on the display.
3. Press in order
M+
,
U
and
TARE
keys to enter technical
parameter setting mode.
4. Press TARE key to select parameter, press
ZERO
key
to enter value and U key to exit.
5. It may be necessary to use
TARE
key to set a value
or use
M+
and
PT
keys to move the active digit.
CALIBRATION
1. Enter Technical Parameter Setting mode (see
Enter Technical Parameter Mode) until display shows
"P1 rEF".
2. Press the
TARE
key until display shows "P2 CAL".
Press
ZERO
key to enter and press
TARE
key until
display shows "CAL".
3. Press
ZERO
key twice and display will show "UNLD".
Remove all weight from the forks. After STABLE
indicator is visible, press the
ZERO
key.
4. The display shows the calibration weight value. Use
the
M+
,
PT
and
TARE
keys to set the weight value.
NOTE: Use M+/PT keys to move active digit, use
TARE key to change value. You can also press
G/N to make value zero.
5. After weight is keyed, press
ZERO
key to enter.
Display will show "LOAD".
6. Add calibration weight onto forks. After STABLE
indicator is visible press
ZERO
key to enter.
Display returns to normal once calibration is successful.
If an error message is shown, try the calibration steps
again; a disturbance may have prevented a successful
calibration. If the problem persists, please contact
Uline Customer Service at 1-800-295-5510.